64 . TGA 8000 Installation and Hardware Guide
Furnace Tube Maintenance
Maintenance of the furnace tube consists of the following procedures:
•
•
•
Removing the Furnace Tube
1.
Make sure that the furnace is in the lowered position.
2.
Remove the exhaust line from the furnace tube side arm.
3.
Remove the clamp around the base of the furnace tube.
4.
Remove the glassware.
CAUTION:
Do not pull the furnace tube very quickly or at an angle to avoid
breaking the thin stem on the furnace.
ATTENTION:
Ne pas tirer le tube du four très rapidement ou à angle pour éviter
de briser la tige mince sur le four.
Replacing the Furnace Tube
This procedure is performed after removing the furnace tube.
1.
Carefully place the tube over the furnace and place on the O-ring at the top of the
metal furnace base.
2.
Position the furnace tube so that the exhaust tube connector is left of center while
the furnace is in the lowered position. The exhaust tube connector should be able to
go into the cool position without hitting the side of the analyzer.
3.
Place the clamp around the furnace tube bottom and the top of the furnace base and
tighten. Make sure that the red piece is just to the left of the exhaust tube.
4.
Replace the exhaust tube to the furnace side arm.
Содержание TGA 8000
Страница 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Страница 5: ...5 Safety and Regulatory Information...
Страница 19: ...19 Prepare the Laboratory...
Страница 25: ...25 Installing the TGA 8000...
Страница 55: ...55 TGA 8000 Hardware Maintenance...
Страница 68: ...68 TGA 8000 Installation and Hardware Guide...
Страница 85: ...85 TGA 8000 Autosampler...