AS 6000 Autosampler . 85
Error Messages from the TGA 4000
Error message ID
Error
Possible Cause Corrective
Action
ERR_DUE_TO_AUTOSAMPLER
AS 6000 entered
fatal error state.
This caused TGA
to enter fatal
error state
Refer to AS 6000
error message
Refer to AS 6000
troubleshooting
(above)
ERR_AUTOSAMPLER_RESPONS Timeout on
command.
Response from
command
transmitted to
AS 6000
I2C connection
cable defective
Replace I2C cable
ERR_AUTOSAMPLER_I2C_WD
No link between
TGA and AS 6000
via I2C
I2C connection
cable defective
Replace I2C cable