![PerkinElmer TGA 4000 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/perkinelmer/tga-4000/tga-4000_installation-and-hardware-manual_1530261039.webp)
Install a TGA 4000 . 39
1.
Remove the fuse holder with a screw driver by gently prying it out of the compartment.
2.
Once the holder is out far enough to hold onto, slide the fuse holder out of the slot.
3.
Remove the fuse (3.15 A) from the 220 V position. This can be done using the point of
a pencil to gently lift the fuse out of the holder.
NOTE: It may be necessary to move the fuse slightly towards the back of the fuse pack
before prying it out of the holder.
NOTE: In the figure above, we are showing you where the fuses go. The fuse that is
installed is for 220V.
4.
Insert the fuse (6.3 A) for 110 V (it goes into the right side when 110–120 V is in the
upright (legible) position).
NOTE: Ensure that the head of the fuse makes contact at both ends of the fuse holder.