Maintenance and Troubleshooting . 97
7.
Select
Include this location in the search
, click
Browse,
and then navigate to
C:\Program Files\PerkinElmer\ServiceIR\Raman.
These options may be displayed automatically as you work through the wizard.
8.
Click
Next
.
The Completing the Found New Hardware Wizard is displayed.
9.
Click
Finish
.
The Found New Hardware Wizard closes.
Problem:
A motorized sample accessory does not go to the correct position
when asked.
Description This means that for some reason the motorized stage is jammed. This
typically happens if the sample holder on the stage is different to that
selected on the Sampling tab in the Scan and Instrument Setup dialog.
Solution 1
Go to the Sampling tab and ensure that the selected sample holder is
correct. If it is not then select the appropriate sample holder.
Содержание RamanMicro 300 Series
Страница 1: ...RamanMicro 300 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanMicro 300 Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 21: ...Raman History Theory and Instrumentation...
Страница 29: ...Preparing Your Facility...
Страница 35: ...Installing Your RamanMicro 300...
Страница 68: ...68 RamanMicro 300 Getting Started Guide...
Страница 69: ...Using the RamanMicro 300...
Страница 90: ...90 RamanMicro 300 Getting Started Guide...
Страница 91: ...Maintenance and Troubleshooting...
Страница 102: ...102 RamanMicro 300 Getting Started Guide...