5
FOOD-6006-01
POWER
The power button should be used to power down the device, and if there is a need for a
hard shutdown in the event the unit freezes and becomes unresponsive . Alternately, use
the Total Power Down method by following on-screen
Menu
Power
Power Off
to
shut down the reader safely .
HOME
The
Home
button can be used to return to the Main menu from any point of the testing
routine .
MENU
The
Menu
button can be used to access the user menu from the main screen .
KEYBOARD
The QuickSTAR
™
Horizon
™
strip reader interface features an onscreen keyboard, allowing
the user to input data into the device by touching the desired letters and numbers .
Select
to temporarily use uppercase letters for a single letter .
Select
&123
to alternate between letters and symbols .
THEORY OF OPERATION
Inserting an adapter containing a lateral flow test strip into the QuickSTAR
™
Horizon
™
strip
reader will initiate a test imaging and analysis sequence in which the reader performs an
optical image scan of the test strip’s results area .
For all tests, it is important to properly insert the test strip into the cartridge adapter, as
well as to properly insert the cartridge adapter into the reader .
Refer to the figures on this page to acquaint yourself with the correct procedure .
Insert the disposable test strip, sample application pad first, into the supplied cartridge
adapter, ensuring that the test strip remains flat against the adapter and that the cartridge
is inserted fully into the reader .