Description . 35
Specification
Temperature Range
0 – 100
o
C
Temperature Readout Resolution
0.1
o
C
Temperature Readout Stability
+ 0.1
o
C
Alarms (Cell holder)
Alarms to be set for high and low
temperature limits through keyboard.
Configuration parameters P-01 and
P-02 (see page 47).
Alarms (Whole system)
If the temperature of the heat sink on
the cell holder reaches 70
o
C the
system is automatically switched off
Accuracy of the set temperature
Better than +0.7
o
C at 0
o
C
Better than + 0.3
o
C at 30
o
C
Better than -1.5
o
C at 100
o
C
Difference between set and actual
temperature in the cell
Better than +1.2
o
C at 0
o
C
Better than + 0.5
o
C at 30
o
C
Better than -3
o
C at 100
o
C
Achievable temperature programming
rate (heating and cooling)
Better than 10 °Cmin
-1
Memory for patterns (non volatile)
2 methods (8 steps)
Stirring Speed
Up to approximately 1800 rpm
Maximum temperature difference from
cell to cell (PTP-6 and PTP-6+6)
Better than 0.8
o
C
at 5
o
C
Better than 0.4
o
C
at 30
o
C
Better than 0.8
o
C
at 70
o
C
Maximum RMS current input
550 mA
Содержание PTP-1
Страница 1: ...Peltier Temperature Programmer User s Guide...
Страница 5: ...Introduction...
Страница 10: ...10 Peltier Temperature Programmer User s Guide...
Страница 11: ...Warnings and Safety Information...
Страница 18: ...18 Peltier Temperature Programmer User s Guide...
Страница 19: ...Installation...
Страница 30: ...30 Peltier Temperature Programmer User s Guide...
Страница 31: ...Description...
Страница 36: ...36 Peltier Temperature Programmer User s Guide...
Страница 37: ...Operation...
Страница 47: ...Configuration and Maintenance...