Troubleshooting

Error

Cause

Possible Solution

Unclosed Lid

The lid is not latched 

properly

1. Check that nothing prevents the lid from closing.
2. Press lid and ensure the handle is locked properly.

Detection system is faulty

1. Turn off the unit.
2. Contact technical assistance.

Fan not working

No Power

1. Check main voltage

2. Check that the voltage of  the unit matches 

     that being delivered by the main power supply.

3. Make sure that the unit is plugged in properly.

4. Check and replace fuses if blown.

Faulty Fan

1. Turn off the unit.

2. Contact technical support.

Display screen shows 

incomplete or 

unusual characters

CPU needs to be reset

1. Turn off the unit. 
2. Wait 10 seconds. 

3. Turn the unit back on.

No display on the screen

No Power

1. Check main voltage.

2. Check that the voltage of the unit matches 

that being delivered by the main power supply.

3. Make sure the unit is plugged in properly.

4. Check and replace fuses if blown.

Faulty display screen

1. Turn off the unit.

2. Contact technical support.

One or several tubes are 

not sealed tightly

The cap is not 

properly sealed or the tube 

is faulty

If dangerous or potentially dangerous sample 

is contained in the tube, apply the proper 

decontamination procedure.

Unit powers on with 

a loud “BANG”

Motor control device 

damaged due to incorrect 

voltage selection

Return unit to OMNI International for 

replacement of motor control device.

Unit powers on but 

motor does not turn

Incorrect voltage selected

Power off the unit and select appropriate 

voltage. If problem persists, contact OMNI 

International for assistance.

Fingerplate lock is 

difficult to disengage.

Knob position

Turn knob slightly counter-clockwise to release 

tension.

Tube leakage

Speed is too high

High temperature or 

pressure inside tubes

Reduce processing time or speed

DO NOT: attempt to service the Bead Ruptor Elite

TM

 Bead Mill in a manner other than 

those discussed in this manual. For any issue that is unsuccessfully corrected using this 
guide, please contact your authorized dealer or call OMNI International at 1-800-776-4431.

Содержание OMNI Bead Ruptor Elite

Страница 1: ...Bead Ruptor EliteTM BEAD RUPTOR ELITE Bead Mill Homogenizer Bead Mill Homogenizer User Manual ...

Страница 2: ... OF MERCHANTABILITY OR OTHERWISE EXPRESSED OR IMPLIED IN FACT OR BY LAW AND STATE OUR ENTIRE AND EXCLUSIVE LIABILITY AND BUYERS EXCLUSIVE REMEDY FOR ANY CLAIM OF DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING OF GOODS OR PARTS THEIR DESIGN SUITABILITY FOR USE INSTALLATION OR OPERATION OMNI INTERNATIONAL WILL IN NO EVENT BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER AND THEIR LIA...

Страница 3: ...lbs The dimensions of the system are withlidclosed 38 1x43 18x33cm 15 x17 x13 withlidOpen 39 37x44 45x61 6cm 15 5 x17 5 x24 25 Operating Environment 4 C to 40 C 39 F to 104 F Humidity 5 to 95 RH Ambient Air The system must be located in an area where the ambient air is clean No emission of solid particles or smoke in the air by adjacent equipment is allowed The level of dust should be comparable t...

Страница 4: ...impaired FCC This device complies with part 15 of the FCC United States Federal Communications Commission Rules Operation is subject to the following two conditions This device may not cause harmful interference and This device must accept any interference received including interference that may cause undesired operation CE This device complies with all CE rules and requirements Changes or modifi...

Страница 5: ...d Port 220 volts Remove Warning stickers from the back of the unit and check that the voltage matches the voltage delivered by your electric power supply CAUTION DO NOT connect the unit to the main power supply until correct voltage is selected Plug the unit into power supply using the supplied power cord CAUTION If the voltage supplied to the unit is higher than what is indicated on the voltage s...

Страница 6: ...must be spaced evenly Place Tube Carriage sold separately on Drive Hub Place Finger Plate on Drive Hub Screw down to secure tubes Close the lid until the locking handle clicks into place CAUTION The unit will not run without the lid fully closed WARNING Do not run the Bead Ruptor EliteTM Bead Mill without a tube carriage installed 2 4 5 1 3 ...

Страница 7: ...on Getting Started Language Setting 1 Menu 2 Home 3 Search 4 Configure 1 2 3 4 Select Menu Select Configure Select Language v wTo navigate through menu Choose from French English German Italian or Spanish Select Home ...

Страница 8: ...o Create new program 1 Menu 2 Program Name 3 Tube Setting 4 Speed 5 Time 6 Processing chamber temperature 7 Run Pause 8 Tube Speed Limit 9 Number of Cycles 10 Time interval or Dwell between cycles 1 2 3 4 5 6 7 8 9 10 00 30 00 30 02 4 50 Max Tube Speed 08 46 2 30 ml Tube Bacteria 30 mL 28 C Ready ...

Страница 9: ...ram with new name Delete program Edit Title Change program name Parameter Operating Range S Speed Energy From 0 8 m s to 8 m s T Time From 0 01 to 9 59 in increments of 1 second C Number of Cycles 1 to 10 D Dwell Pause between Runs 0 00 to 9 59 Run time indicator Pause Run button Cycle indicator Stop run button Bacteria 30 mL 28 C 02 35 ...

Страница 10: ...n Faulty display screen 1 Turn off the unit 2 Contact technical support One or several tubes are not sealed tightly The cap is not properly sealed or the tube is faulty If dangerous or potentially dangerous sample is contained in the tube apply the proper decontamination procedure Unit powers on with a loud BANG Motor control device damaged due to incorrect voltage selection Return unit to OMNI In...

Страница 11: ......

Страница 12: ...PN 03 241E Rev C 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com ...

Отзывы: