Put the new lid in place and attach loosely with the hex screws. Ensure that the lid is centered.
Press it against the cooling plate surface and tighten the hex screws.
Decontamination information
The following instructions should be followed before performing any service or disassembly with the Migele Gel
Electrophoresis Unit.
Note:
Use gloves when performing any service with parts that may have been into contact with patient samples (blood
spots).
Clean the instrument surfaces with deionized water or with a 70%-80% ethanol solution. Do NOT use any other liquid
for cleaning.
Decontamination information form
Product type ________________
Serial no ________________
This is to inform those who maintain, repair or transfer the product that the instrument and its subunits have been
treated according to the procedure given above.
Date ________________
Signature _________________________________
This procedure is also to be used in order to eliminate or reduce hazards when involved in removal from use,
transportation or disposal.
Copy this page, fill in the data and include the page with the instrument.
2118-0010 - Migele
®
Gel Electrophoresis Unit User Manual
Maintenance
28