Description of parts
Two to four 2118-0010 Migele Gel Electrophoresis Units can be attached to series with one circulating water bath (for
example, 1023-0010, 1023-0020 or equivalent).
Two to four 2118-0010 Migele Gel Electrophoresis Units can be attached to one programmable power supply (e.g.
CWP-2000 or equivalent) depending on the model of the power supply.
Components in the system
This section describes the components in the Migele Gel Electrophoresis Unit system.
Cooling plate
Ceramic (aluminium oxide) cooling plate measures 260 x 260 mm, supports the gel, and provides uniform temperature
control. Aluminium oxide is an excellent heat conductor and electrical insulating material.
To facilitate the correct positioning of electrophoresis gels, the surface of the cooling plate is screened with a grid
measuring 240 x 240 mm.
Figure 3: Grid
Warning:
Do not use the Migele Gel Electrophoresis Unit at above 2000 V.
Warning:
Use water in the circulating water bath. The maximum pressure is 0.5 bar.
2118-0010 - Migele
®
Gel Electrophoresis Unit User Manual
Description of parts
8