Safety Practices. 29
Fuses
WARNING
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
The fuse is located at the rear of the instrument.
1.
Turn off and unplug the instrument.
2.
Carefully open the compartment latch where the fuse is located.
3.
Disconnect the fuse.
4.
Replace with a new T2.0AL fuse. One spare is contained in the power module.
5.
Close the compartment door.
6.
Plug in the instrument and turn on.
Location of latch on the
fuse compartment door