Fitting and Removing the Optical Bench
In the first instance, the General Purpose Optical Bench is installed by a PerkinElmer Service
Engineer, who will install: a right-hand beam moveable mirror and any other internal parts
required; a KBr window; and an external accessory bracket in the handhold on the right of the
instrument (if not already fitted). This external accessory bracket should not be removed.
Figure 2 Handhold with external accessory bracket fitted
The optical bench should not be removed unless:
•
You want to use the external beam port on the right of the instrument with a different
external accessory;
•
The spectrometer must be moved to a new location.
For information about moving your spectrometer, refer to the
Spectrum 100 Series User’s
Guide
(L1050021) or
Spectrum 400 Series User’s Guide
(L1050056). These are distributed
as .pdf files on the
Spectrum Manuals CD
(L1050002).
Before you begin, make sure you have read and understood the “Warnings and Safety
Information” beginning on page 2.
Fitting a Window in the Right-hand External Beam Port
1.
Switch off the spectrometer, and then remove the power cable.
2.
Undo the locks on the front of the spectrometer, and then open the main cover
.
For detailed instructions, refer to the
Advanced Maintenance
section of the
Spectrum 100
Series User’s Guide
(L1050021) or
Spectrum 400 Series User’s Guide
(L1050056). These
are distributed as .pdf files on the
Spectrum Manuals CD
(L1050002).
3.
Using the hexagonal wrench supplied, slacken and remove the M2.5 screws securing the
right-hand external beam port cover to the main cover.
4.
Remove the beam port cover and seal. Retain for further use.