L1050070A
10
Mirror module
Bolt the two pillars and cross-bar that support the parabolic mirror to the optical bench,
oriented to that the mirror faces the detector. See Figure 6 for details of fixing position.
Figure 10 Mirror assembly fitted to bench
Later, you can adjust the vertical alignment of the parabolic mirror to the detector using the
adjustment screw in the cross bar. See “Optimizing Performance” on page 16.
Configuring Spectrum or Spectrum Express Softw are to Use
an External Detector
The General Purpose Optical Bench with External LiTaO
3
Detector accessory is supported by
Spectrum version 6.3.5 (or later) for the Spectrum 400 Series Spectrometers, or Spectrum
version 6.3.3 (or later), or Spectrum Express version 1.2 (or later), for the Spectrum 100 Series
Spectrometers.
This section describes how to create a modified instrument configuration file for use with an
external detector.
NOTE: It is not necessary to power the instrument to complete this procedure.
1.
Browse to C:\pel_apps\bin\
The instrument configuration file is named C*****.cfg, where ***** is the five-digit serial
number of the instrument.
The spectrometer serial number is located on its base casting, in the sample compartment,
under the sample accessory baseplate.
2.
Copy the appropriate C*****.cfg file and save it as, for example, C*****.intdet
NOTE: Make sure you save this backup of the original instrument configuration file. If you
remove the optical bench, you will re-instate this file.
M6 fixing bolt
M6 fixing bolt
Vertical alignment
adjustment screw