64 . Frontier IR Dual-Range Systems User's Guide
The Desiccant Indicator in Detail
The optical system of the spectrometer is purged at the factory. This protects the sample
compartment windows, external port windows, or beamsplitters from being damaged by
humidity. Replaceable packs of desiccant maintain the purge.
The top panel of the instrument includes a desiccant indicator (Figure 28), whose sectors
change sequentially from blue to pink as the desiccant becomes exhausted. Change the
desiccant packs in the instrument when the sector marked 10 becomes pink, but while the
sectors marked 15 and 20 are still blue. These numbers correspond to the approximate
% Relative Humidity in the instrument.
NOTE: A similar desiccant indicator is fitted in the desiccated compartment (Figure 34 on
page 68). This is used for storing sample compartment windows when not in use.
Figure 28 The desiccant indicator: when desiccant packs should be changed
CAUTION
If all three sectors of the desiccant indicator are pink, then you must
change the desiccant.
The instrument optics may be fogged. Do not switch the instrument
either ON or OFF until all sectors are blue.
Sector
Action Required
All sectors blue
None.
The humidity levels in the instrument are below approximately 10%
Relative Humidity.
Sector 10 pink
Desiccant change is recommended.
If the instrument has been switched OFF for an extended period, do not
switch ON until you have changed the desiccant and all sectors are blue.
Sector 10 & 15
pink
Replace desiccant the immediately to avoid instrument damage.
Do not switch the instrument ON or OFF until you have changed the
desiccant and all sectors are blue.
Pink
Blue
Blue
Содержание Frontier IR
Страница 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 Frontier IR Dual Range Systems User s Guide...
Страница 11: ...Warnings and Safety Information...
Страница 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Страница 34: ...34 Frontier IR Dual Range Systems User s Guide...
Страница 35: ...Unpacking and Installation...
Страница 45: ...Using the Spectrometer with Spectrum...
Страница 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Страница 57: ...Routine Maintenance...
Страница 78: ...78 Frontier IR Dual Range Systems User s Guide...
Страница 79: ...Advanced Maintenance...
Страница 94: ...94 Frontier IR Dual Range Systems User s Guide...
Страница 95: ...Appendices...