Routine maintenance
54
Reverse the procedure to reinstall the magazine table and magazines.
Changing filter slides
1.
Lift the instrument lid.
2.
You will see the ends of the excitation and emission filter slides directly
in front of you. Pull out the slide you want to get access to.
3.
Add, remove or exchange filter modules as needed.
4.
The filter module snaps into place in the filter slide.