40 . DSC 4000 Installation and Hardware Guide
1.
Remove the fuse holder with a screw driver by gently prying it out of the compartment.
2.
Once the holder is out far enough to hold onto, slide the fuse holder out of the slot.
3.
Remove the fuse (3.15 A) from the 220 V position. This can be done using the point of
a pencil to gently lift the fuse out of the holder.
NOTE: It may be necessary to move the fuse slightly towards the back of the fuse pack
before prying it out of the holder.
NOTE: In the figure above, the fuse that is installed is for 220 V.
4.
Insert the fuse (6.3 A) for 110 V (it goes into the right side when 110–120 V is in the
upright (legible) position).
NOTE: Ensure that the head of the fuse makes contact at both ends of the fuse holder.