XRpad2 4336
For Your Safety
Automatic Exposure Detection (AED) Mode
If a Problem Occurs
WARNING
The AED mode requires a sufficient X-ray dose rate to the detector
surface. The required dose rate can vary between X-Ray tube,
generator, and X-ray voltage. The system integrator must evaluate the
AED operation with the complete X-ray setup in order to secure a
proper image acquisition in AED mode.
WARNING
The external sync mode is the default acquisition mode, and the
detector needs to switch into the AED mode to use the Auto Exposure
detection. It is important that a settling time for the AED is
implemented by the system integrator. If the settling time is too short,
the image can show artifacts.
WARNING
The AED on-time is limited and must be turned off directly after
acquiring the image. Staying longer in AED mode will reduce the
battery duty cycle and heat up the detector.
WARNING
Do not apply any, handling, loading, mechanical shock, or electronic
noise to the detector while it is in AED mode. These actions can start
an unwanted acquisition (false trigger). If a false trigger is applied to
the detector, the detector will not be able to react on a real X-ray
exposure until it has acquired the false trigger image.
WARNING
If any abnormal condition, such as smoke, fumes, or strange sounds,
is evident, turn off the X-ray detector, turn off and unplug the power
supply from the AC outlet, and immediately ask your establishment’s
safety representative to contact your dealer, distributor, or device
manufacturer.
Further use under abnormal conditions may result in severe personal
injury, death, or substantial product damage.
WARNING
When liquid has been spilled into or on any part of the X-ray detector
or power supply (if applicable), or when the X-ray detector, its
component, or accessory is dropped, unplug the power supply from
the AC outlet, and immediately ask your establishment’s safety
representative to contact your dealer, distributor, or device
manufacturer.
Further use under abnormal conditions may result in severe personal
injury, death, or substantial product damage.
Содержание XRpad2 4336
Страница 14: ...This page intentionally left blank xiv www perkinelmer com 68153 01 List of Figures XRpad2 4336 ...
Страница 16: ...xvi www perkinelmer com 68153 01 List of Tables XRpad2 4336 This page intentionally left blank ...
Страница 62: ...This page intentionally left blank XRpad2 4336 46 68153 01 ...
Страница 66: ...This page intentionally left blank 50 www perkinelmer com 68153 01 Index XRpad2 4336 ...