XRpad2 4336
20
68153 01
D
IGIT
A
L
I
MAGING
includes the specification for the XRpad IPU-2.
Table 11
Specification of the XRpad IPU-2
Electrical Specification
AC Input Voltage [16]
100 V … 240 V
AC Frequency [16]
50 Hz/60 Hz
DC Output [8]
12.5 V/5 A, 15 V/1 A
(voltage level is dependent of load)
Trigger In Signal from Hand Switch [6]
5 V … 24 V/10 mA (SELV)
Trigger Out Signal to Generator [5]
Same level as Trigger In Signal
Trigger In Signal [7, 9]
3.3 V … 5 V (SELV)
Trigger Out Signal [7, 10]
3.3 V
DC Output 5PF [7]
5 V/100 mA
Mechanical Specification
Size
260 mm (L) × 205 mm (W) × 50 mm (H)
Temperature Ranges
Operating
10°C to 35°C
Transportation/Storage
–10°C to 70°C
Relative Humidity
Operating
10% to 90%
Transportation/Storage
0% to 90%
Ingress Protection Rating
IP40 rated (protection against particles > 1 mm)
WARNING
All external signals that are connected to the IPU-2 (especially PREP/
EXPOSE and Trigger signals) should be from a Separated or Safety
Extra-Low Voltage (SELV) circuit. Ignoring this warning may result in
electric shock, which may result in severe personal injury, death, or
substantial product damage.