5.
Press the ON/OFF key to turn on the control unit. When the stirrer speed is displayed, adjust the
desired stirrer speed by using the MINUS-key or PLUS-key.
Figure 44 Pressing MINUS-key or PLUS-key
NOTE:
When you press the mode key, power adjustment mode is displayed. The power settings should be set to
P 25.
NOTE:
The default power set is 50. Whenever turning the main switch on or off, you should set the power to 25.
Figure 45 Power set to 25
Connect to the 8 Cell Water Jacketed Cell Changer
1.
Connect two stirrer control cables to the 4-pin socket of the distriBOX4.
Figure 46 Connecting the two stirrer control cables