CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
28
PERISO SA
properly plugged.
Check that the emergency button it
is not pressed.
The arm does not work.
The motor plug of the arm is
regularly pushed forward and well
connected, on the back of the
power module?
The plug of the control shelf is
regularly pushed forward, on the
back of the power module?
The CTU Mega 20 is connected to
a 230V 50 / 60 Hz power source?
The key has been turned to the ON
position?
Check that it is not pressed the
emergency button.
Visualization
With the device on you cannot select some
or all items from the display menu when
entering the work templates.
Is the cable between the device
controller and the power module is
connected properly?
Check the display of the
same functionality in
different points.
Ensure that the display is
not greasy or dirty and that
there is no cover film.
The device turns on but the display is not
clearly visible.
Is the cable between the device
controller and the power module is
connected properly?
Try adjusting the contrast
of the display using the
trimmer on the back of the
control module by turning
the same, very slowly, first
to the right and then to the
left.
When the device is on, the words written on
the screen disappears.
Make sure the USB cable between
the control module and the power
module is connected correctly.