background image

5.1 CORRECT MEASUREMENT METHOD

The temperature sensor is located at the end of the metal sheath of 

the sheath type TEMPERATURE PROBE. To accurately test internal 

temperature insert the probe into the item you want to measure to a 

distance at least 15 times the diameter of the sheath

5.2 CONNECTOR CONFIGURATION

 

 

5. TEMPERATURE MEASUREMENT

P.10

Содержание TC0376

Страница 1: ...Datalogger Thermometer TC0376 Instruction Manual...

Страница 2: ...electing the Temperature Scale 4 4 Data Hold Operation 4 5 Back light Operation 4 6 Relative Operation 4 7 Record Operation 4 8 MAX MIN Operation 4 9 Auto Power Off 4 10 Low Battery Condition 5 Temper...

Страница 3: ...nsor NUMERICAL DISPLAY 5 digital liquid crystal displays MEASUREMENT RANGE 100 C 400 C 148 F 752 F RESOLUTION 0 01 C 0 02 F SENSOR TYPES Platinum resistance temperature sensor for pt 100 4 wires ALPHA...

Страница 4: ...28 C add the following tolerance into the accuracy spec 0 005 of reading 0 01 C 0 005 of reading 0 02 F NOTE The basic accuracy Specification does not include the error of the probe please refer to th...

Страница 5: ...TERY LIFE Approx 100hours STANDARD ACCESSORY Pt 100 Probe class A Instruction manual 9V Battery windows software carrying case USB cable OPTIONAL ACCESSORIES AC adapter RS 232 Cable P 4 2 SPECIFICATIO...

Страница 6: ...yed The Maximum value is now being displayed This indicates that the display data is being hold This indicates Auto Power Off is enabled Readings are being logged The reading is now under Relative Mod...

Страница 7: ...ure sensor connector LCD display Power ON OFF and Back light button Relative readout button REC button MAX MIN control button 8 9 10 11 12 7 HOLD button C F control button PC Interface External DC 9V...

Страница 8: ...he default scale setting is set at Celsius C scale The user may change it to Fahrenheit F by pressing button and vice versa to Celsius and change default scale 4 4 DATA HOLD OPERATION The user may hol...

Страница 9: ...ATION Starting and Stopping Logging Setup memory clear and PC communications are inaccessible during logging Set the logging interval See Changing the Logging Interval Press button to start logging Th...

Страница 10: ...mode one may press and hold button for two seconds 4 9 AUTO POWER OFF By default when the meter is powered on it is under auto power off mode The meter will power itself off after 30 minutes if no ke...

Страница 11: ...end of the metal sheath of the sheath type TEMPERATURE PROBE To accurately test internal temperature insert the probe into the item you want to measure to a distance at least 15 times the diameter of...

Страница 12: ...th and then press button to select The right two digits blink 6 5 Press or button until the display shows the correct day and then press button to select The left two digits blink 6 6 Press or button...

Страница 13: ...mode 7 2 Press button to enter Interval setting mode The left two digits blink 7 3 Press or button until the display shows the correct hour 24 hour format and then press button to select The right two...

Страница 14: ...gram shows each button function when user enter into calibration mode NOTE Turn the Power OFF before attempting following SETUP Setup mode is canceled during below procedure if POWER button is pressed...

Страница 15: ...uring the process press button to exit calibration mode 9 2 RECALL DEFAULT FACTORY SETTING VALUE Press and hold buttons and then power on the meter to enter setup mode Press button to enter calibratio...

Страница 16: ...0MHz or higher 128 MB RAM At least 50 MB hard disk space available to install SE376 Recommended screen resolution 1024X768 or above 4 INSTALL SE 376 AND USB 120 CONNECTOR DRIVER 1 We recommend close a...

Страница 17: ...SE376 software automatically it will copy SE376 exe executable file and help file to your hard disk default is c program files SE376 9 After finish installation plug USB 120 connector to PC usb port...

Страница 18: ...COMx is in the list the connector is ready to work Open Retrieve files from the disk Save Save the present data to the disk Print Send data in the present table to the printer Pause Stop data recordin...

Страница 19: ...nge will be effective immediately It will keep up to 5000 data points when data points exceed this limit the oldest 100 points will be replaced sequentially You can Zoom this graph by using mouse To Z...

Страница 20: ...have thermometer connected to computer select Load Data to start loading the recorded data from the thermometer There will be a progress indicator to show the loading progress If error occurs just cli...

Страница 21: ...ata set to show its graph and tabular on the right side window You can choose a rectangle area on the graph to zoom in for detail The vertical dash line in the graph will match the highlight row in th...

Страница 22: ...he decimal separator is comma in your area you cannot use CSV file because CSV file use comma to separate data This will cause confusion for example 23 6 C 24 6 C However you can use TXT files instead...

Страница 23: ...icator will appear Replace a 9 Volt battery immediately to ensure the unit functions properly 11 3 AC ADAPTER CONNECTION When the AC adapter is used insert the plugs of the adapter into the DC9V conne...

Страница 24: ...static and dry cleaning cloth P 23 12 MAINTENANCE Do not use cleaning agents that contain carbon or benzenes alcohol or anything similar to clean the product since these substances damage the surface...

Страница 25: ...Accuracy Measurement Range Temperature Sensor Dimensions Temperature Sensor Length Cable Length Platinum resistance thermometer sensor Pt 100 4 wires IEC751 class A 0 15 C 0 002t t measurement temper...

Страница 26: ...r products and services please send us an email cs perfectprime com For B2B or project based application please send an email sales perfectprime com FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISI...

Страница 27: ...HORMES LIMITED cs perfectprime com G F UNIT 3 61 GLENTHORNE ROAD LONDON W6 0LJ UNITED KINGDOM 44 203 7695377 Retailer Email Address Telephone...

Отзывы: