Instruction Manual peqSTAR X thermal cyclers
PEQLAB_v1212E
15
9.2.4.5
Gradient Control
Using the Checkbox
you can
decide whether in a gradient step the plateaus (Plateaus
synchronized) or the Ramps (Ramps synchronized) should
be synchronized for all rows.
9.2.4.6
Tube Control
With the Checkbox
you have the
possibility to adjust temperature regulation for a
temperature step accordingly to sample volume. If
necessary, the temperature regulation will perform short-
term deliberate overshoots and undershoots depending
on the entered sample volume during PCR run.
9.2.4.7
Emulation
Using the Checkbox
the thermal cycler
will perform PCR like the chosen cycler, emulating its
performance. This function could be useful if a PCR
program should be transferred to the peqSTAR from
another type and model of thermal cycler.