(2) REPROCESSING SYSTEM
To reprocess a PENTAX single-channel videoscope, first an enzymatic detergent and then a high-level disinfectant
or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and scope components
(air/water valve, suction valve, etc.). Exposure times of detergent and disinfectant/sterilant must be strictly adhered
to as recommended by each respective manufacuturer.
Please note that all solution entrance ports and flow pathways are illustrated above.
CAUTION:
BEFORE IMMERSING:
a) The ‘Red’ ETO gas sterilization venting cap must be taken OFF.
b) The PVE Soaking cap should be securely ON the electrical contacts.
NOTE:
Prior to exposure of all internal channels to an enzymatic detergent and high-level disinfectant (or steri-
lant), all brushable channels should be manually cleaned with cleaning brushes.
EG-290Kp/EC-380Kp
W
a
ter Feeding Ch
a
nnel
R
ubb
er Inlet
S
e
a
l
Ch
a
nnel Inlet
A/W In
s
tr
u
ment Ch
a
nnel Cle
a
ning Ad
a
pter OF-B115
Air/W
a
ter Ch
a
nnel
Cle
a
ning Ad
a
pter OF-G17
PVE
S
o
a
king C
a
p
(ATTACH)
ETO C
a
p
(REMOVE)
Air Feeding T
ub
e
W
a
ter Feeding T
ub
e
Su
ction T
ub
e
Su
ction Nipple
L
u
er-
S
lip
S
yringe
with
Cle
a
ning/Di
s
infectiing
S
ol
u
tion
L
u
er-
S
lip
S
yringe
with
Cle
a
ning/Di
s
infectiing
S
ol
u
tion
Air Feeding Ch
a
nnel
In
s
tr
u
ment Ch
a
nnel
Air/W
a
ter Nozzle
Содержание EC-380LKp
Страница 1: ...PENTAX VIDEO GI SCOPES EG 290Kp OWNER S MANUAL EC 380LKp...
Страница 50: ...47 MEMO...
Страница 51: ...48 MEMO...