HAFFMANS ICC
Page 32 of 49
ICC
9.3
IAM maintenance (weekly)
Procedure:
1. Disassemble the Teflon sealing tap (pos. 112),
2. Rinse the sealing tap in clean water,
3. Lubricate the sealing tap with silicon grease,
4. Assemble the sealing tap in the IAM (fit the red disc first, followed by the O-ring and
then the nut).
9.4
ICC-2000 cleaning
It is recommended to clean the ICC-2000 daily if measurements have been carried out.
Method:
1. Connect a water hose to the pressure valve (pos. 57),
2. Place a cup under the ICC alternatively, connect a hose to the needle. If you connect a
hose here, remove the rubber seal (pos. 36) first,
3. Open the needle valve (pos. 13),
4. Slowly open the water tap and let the water run for a few minutes,
5. Open and close the needle valve (pos. 13) regularly (the instrument will now be
completely cleaned internally,
6. Shut the water tap and disconnect the hose (s).
9.5
IAM cleaning
It is recommended to clean the IAM-2000 daily with water if measurements have been carried
out.
Warning
Used lye should be disposed of according to the national
legal environmental and health regulations.
Содержание 110.000
Страница 1: ...HAFFMANS FOODANDBEVERAGE PENTAIR COM INSTRUCTION MANUAL INPACK 2000 CO2 CALCULATOR TYPE ICC ...
Страница 2: ......
Страница 3: ......
Страница 5: ......
Страница 11: ...HAFFMANS ICC Page 10 of 49 ICC ...
Страница 13: ...HAFFMANS ICC Page 12 of 49 ICC ...
Страница 17: ...HAFFMANS ICC Page 16 of 49 ICC ...
Страница 23: ...HAFFMANS ICC Page 22 of 49 ICC ...
Страница 31: ...HAFFMANS ICC Page 30 of 49 ICC ...
Страница 35: ...HAFFMANS ICC Page 34 of 49 ICC ...
Страница 37: ...HAFFMANS ICC Page 36 of 49 ICC ...
Страница 41: ...HAFFMANS ICC Page 40 of 49 ICC ...
Страница 43: ...HAFFMANS ICC Page 42 of 49 ICC ...
Страница 47: ...HAFFMANS ICC Page 46 of 49 ICC ...
Страница 48: ...HAFFMANS ICC ICC Page 47 of 49 APPENDIX VII SOFTWARE SUMMARY ...
Страница 49: ...HAFFMANS ICC Page 48 of 49 ICC ...
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......