Error codes
Error code
Description
Solution
2001
Parameter error
Invalid entry, choose valid value
2005
Input value is not valid
Invalid entry, choose value within range
2101
Weigher not stable
Wait for stable and try again
2102
Parameter exceeds maxload
Remove load / edit max load setting
2103
Parameter below zero
Check if scale is blocked
2104
Not in zero range
Remove load
2105
Arithmetic overflow occurred
Change calibration levels
2106
A/D reads all 1’s
Check load cell connection
2107
A/D reads all 0’s
Check load cell connection
2108
Gain ref. < zero ref.
Change calibration levels
2109
Gain > 0.99984741211
Change calibration levels
2110
Save error
Contact PENKO
2111
Flash ROM exhausted
Contact PENKO
2112
Error on header creation
Contact PENKO
2113
Error on date write
Contact PENKO
2114
Header validation failed
Contact PENKO
2115
De-active old data fail
Contact PENKO
2116
Load errors
Contact PENKO
2117
Item not found in store
Contact PENKO
2118
Error in stored data
Contact PENKO
2119
Bad calibration
Change calibration levels
2120
Action not enabled
-
2121
Multi-point not found
Add multi-point calibration
2122
Calibration table full
Remove calibration points
2123
Not allowed, tare active
Deactivate tare
2124
Action not allowed
-
2125
ADC no power
Check power supply
CCCCCC
No proper calibration available
Check calibration setting
UUUUUU
Underflow
Check load cell
Check platform construction
OOOOOO
Overflow
Check load cell
Check platform construction
======
Display overflow; Exceed
maximum display value (max.
load)
Reduce load on platform
Содержание 1020 FMD
Страница 1: ...1020 FMD 1 PENKO Engineering B V Your Partner for Fully Engineered Factory Solutions Manual 1020 FMD...
Страница 7: ...1020 FMD 7 1 Overview Option 1 Option 2...
Страница 19: ...1020 FMD 19 RS232 communication RS422 communication with multiple devices...
Страница 21: ...1020 FMD 21 Hold function Peak Valley and T I R function...
Страница 45: ...1020 FMD 45 Enter the reference weight apply with the Enter button The device is now calibrated...
Страница 69: ...1020 FMD 69 Action Hysteresis Set the hysteresis for each output The hysteresis can be positive or negative...
Страница 99: ...1020 FMD 99...
Страница 103: ...1020 FMD 103...
Страница 106: ...1020 FMD 106 Appendix I Menu structure...
Страница 107: ...1020 FMD 107...
Страница 108: ...1020 FMD 108...