Dose setting
When you have
finished the priming
process, press
to
enter dose setting.
The factory setting
for a dose is 0.5U. The
minimum dosage for
an injection is 0.5U.
Press
or
to adjust the dose. The default dose limit for a
single injection is 60.0U. You can increase the dose limit in the
“Settings Menu”.
Press or
to increase/decrease the number by 0.1U.
Press and hold
or
to increase/decrease the number by 1.
Press
to cancel the dose setting and return to the “Remain.
Ins.” screen.