III. OPERATING PROCEDURES FOR LIQUID STERILIZATION
WARNING: Burns and bodily injury can occur if a sterilizing cycle is interrupted or “aborted” in the follow-
ing manner:
"
Autoclave has been in Sterilize mode for five minutes or more after a cold start.
"
Function control is then turned to Vent and Open Door light is illuminated. If this situation develops,
DO NOT open door even though Open Door light is illuminated. Wait ten minutes before attempting to
open door.
NOTE: Use only Pyrex
®
or stainless steel containers for liquids sterilization. Containers should only be cov-
ered with paper caps.
A. Preparation and Sterilization
1. To sterilize liquids, follow the procedures A-F, Section 3-II “Operating Procedures For Normal Steriliza-
tion.”
B. Vent
1. After the buzzer sounds and Sterile light illuminates, turn the Function control quickly past the Vent
position to Off position.
C. Cool Down
1. Allow autoclave to cool down until pressure gauge indicates “0” (approximately 25 minutes).
D. Unlock Door
1. After ensuring Open Door light is illuminated and pressure gauge indicates “0,” lift door locking handle
while depressing Open Door button on front of Magna-Clave.
E. Drying
1. After unlocking door, allow door to stand partially open with Function control still in the Vent position.
IV. OPERATING GUIDELINES
A. The Magna-Clave may be preheated by turning the Function control to Vent 30 minutes before the
autoclave is needed. This will allow for a faster desired sterilizing temperature.
B. Water should only be added to the reservoir when the Function
control is in the Vent or Power Off position. Demineralized or
distilled water must always be used to prevent corrosion.
The minimum water level should always reach the level
indicating mark on the back of reservoir.
C. Do not attempt to turn the Function control from Sterilize to
Fill or from Vent to Sterilize. There are mechanical stops
which prevent the Function control from being turned in this
manner.
D. Improper sterilization may result if chamber is overloaded or
too crowded.
E. The Time control must be set at the beginning of each cycle
to the desired sterilizing time. If the Time control is not
properly set, the buzzer or chime will sound and the Sterile
light will illuminate. This indicates the end of the sterilizing
cycle, and will not allow completion of the cycle.
NOTE: Do not set the Time control for an interval of less
than five minutes.
SECTION 3
OPERATIONS
TEMPERATURE
GAUGE
INDICATOR
LIGHTS
PRESSURE
GAUGE
TEMPERATURE
CONTROL
DOOR
HANDLE
TIME
CONTROL
FUNCTION
CONTROL