13
SECTION 2
PREPARATIONS FOR STERILIZATION
2. Surgical gloves
NOTE: Disposable gloves should not be sterilized.
a. Clean and dry gloves.
b. Place a square of muslin or other absorbent towel into the glove up to the finger.
c. Place a strip of muslin or other absorbent towel around the cuff and fold it back.
d. Place a process or spore test indicator one glove. Wrap gloves in muslin or other packaging
material. (Refer to Table 2-2.)
e. Place wrapped packs of gloves on end in a perforated tray. Leave space between packs to allow
steam to circulate.
f.
Load trays onto rack in the sterilizer. Leave adequate space between trays for steam circulation.
Ensure that packs do not touch sides of chamber.
CAUTION: Do not overload. Do not seal container. Inadequate sterilization may result.
g. Place pack containing sterilization monitor in the front bottom of the sterilizer.
h. Refer to Section 3 for operating instructions.
IV. LIQUIDS PREPARATION
A. Place liquids in a heat proof glass container, filled 2/3 full.
B. Cover container loosely using a paper or metal cap. Do not seal container.
C. Place container on a perforated tray. Include a process or spore test indicator. Load onto the rack in the
sterilizer. (Refer to Section 3 for operating instructions.)