P/N 82-00021 UM3105KL-AM_3105KG-AM_REV20220519
17
OPERATING INSTRUCTIONS (CONT)
BMI Function With Digital Height Measurement
Note: Digital height measurement is available with the optional digital height rod purchased separately. The
measuring range for digital height data is 3 ft 11.2 in
– 6 ft 7.5 in / 120 cm – 202 cm.
1. Make sure there is nothing on the weighing platform and the
display shows “0.0”.
2. Ask the patient to step on the scale. The scale will begin to
display the patient’s weight and will lock on the patient’s
weight when completed. Depending on the movement of the
patient on the scale, it may take several seconds for the scale
to lock onto the weight.
3. After obtaining the patient
’s weight, and with the patient
remaining on the scale, press .
4. A. If the height rod is set at its home position the display will
read 3 ft 11.2 inches (when weighing in pounds) or 120 cm
(when weighing in kilograms). Move to step 5.
B. If the height rod is not set at its home position, the display
will instruct the user to reset the height rod to its home position
or prompt the user to press an arrow key to enter height
manually. Push the height rod down to its home position.
When at home position the display will read 3 ft 11.2 inches
(when weighing in pounds) or 120 cm (when weighing in
kilograms). Move to step 5.
Note:
Resetting the height rod to
the home position is only needed once after a power up.
There is no need to reset for each height measurement if the
scale has remained on.
5. With the headpiece folded down, slide the headpiece above to
the height of the patient. Then raise the headpiece into its
horizontal position.
Ensure that the patient’s posture is upright,
and that the pati
ent’s head is straight and level. Slide the
headpiece down until it rest
s on the patient’s head. The
display will read t
he patient’s height.
6. Once the height is determined, press .
7. The disp
lay will show the patient’s BMI.
The BMI summary screen will display for approximately 10
seconds or until is pressed to return to normal weighing
mode. After exiting from the BMI summary screen if the patient
is still on the scale the screen will display the weight
measurement again.
BMI
Enter
Zero
(Clear)