background image

Pedigo P-2040-S Blanket Warming Cabinet Operation and Care Manual

5

Error Code

1. Control displays "ooo".

2. Control displays "CCC".

Possible Cause

A. Sensor is open circuited.

B. Associated wiring is open

circuited.

C. Control is faulty.

A. Sensor is short circuited.

B. Associated wiring is short

circuited.

C. Control is faulty.

Action Required

Detach the sensor from the terminal
block.  Use an Ohm meter to measure
the resistance of the sensor.  Check
sensor at 32°F (0°C) using a container
of ice water.  If Ohm reading is 100,
replace display.  If Ohm reading is not
100, replace sensor.

Check wires for integrity.  Check for
proper and secure connections at the
thermostat and terminal block.  If
necessary, re-secure the faulty
connections.

Energize system after the above steps
have been completed.  If control still
reads "ooo", call service technician.

Detach the sensor from the terminal
block.  Use an Ohm meter to measure
the resistance of the sensor.  Check
sensor at 32°F (0°C) using a container
of ice water.  If Ohm reading is 100,
replace display.  If Ohm reading is not
100, replace sensor.

Check wires for integrity.  Check for
proper and secure connections at the
thermostat and terminal block.  If
necessary, re-secure the faulty
connections.

Energize system after the above steps
have been completed.  If control still
reads "CCC", call service technician.

This chart is provided for the assistance of qualified technicians only and is not intended for use by untrained or unauthorized 
service personnel.  If your unit is not operating properly, check the following before calling your authorized service agent.  
Check the power applied to the unit.  Plug in outlet?  Fuse OK?

Do not attempt to repair or service beyond this point.  Contact manufacturer for nearest authorized service agent.  
Repairs made by any other service agent without prior authorization by manufacturer will void the warranty on the unit.

Tr o u b l e   S h o o t i n g   G u i d e

DISCONNECT UNIT 

FROM POWER SOURCE 

BEFORE CLEANING 

OR SERVICING.

EXT WIR

SENSOR

EXT WIR

SENSOR

 SET

prg

!

°

F

ERROR CODE

INDICATOR L.E.D.

HEAT

INDICATOR L.E.D.

ERROR

CODE

DISPLAY

TEMPERATURE

SET

BUTTON

L.E.D.1

INCREASE/DECREASE

ROCKER BUTTON

ooo

POWER
ON/OFF

ROCKER SWITCH

 SET

prg

!

°

F

ERROR CODE

INDICATOR L.E.D.

HEAT

INDICATOR L.E.D.

ERROR

CODE

DISPLAY

TEMPERATURE

SET

BUTTON

L.E.D.1

INCREASE/DECREASE

ROCKER BUTTON

CCC

POWER
ON/OFF

ROCKER SWITCH

Содержание P-2040-S

Страница 1: ...P 2040 S Blanket Warming Cabinet Manual...

Страница 2: ...Heat system incorporating low watt density cable mounted around each compartment This system provides evenly distributed heat without the need of a heat circulating fan Each compartment is controlled...

Страница 3: ...NOT BLOCK SENSOR BY OVERLOADING CABINET WITH BLANKETS OVERHEATING COULD OCCUR THERMOSTAT and INDICATOR L E D SEQUENCE Whenever the thermostat is turned ON the heat indicator L E D will indicate the p...

Страница 4: ...oor gaskets are kept free of any debris that may accumulate Good equipment that is kept clean works better and lasts longer CLEAN THE UNIT REGUL ARLY 1 Disconnect the cabinet from the power source 2 R...

Страница 5: ...heck wires for integrity Check for proper and secure connections at the thermostat and terminal block If necessary re secure the faulty connections Energize system after the above steps have been comp...

Страница 6: ..._____________________ ____________________________________________________________________________ ____________________________________________________________________________ ________________________...

Страница 7: ...Y 2 RL 3736 12 TRANSFORMER 2 TN 3972 13 FUSE 1 4 A SLOW BLOW 2 FU 33332 14 SENSOR SERVICE KIT 2 14785 15 TERMINAL BLOCK SENSOR 2 BK 33546 16 TERMINAL BLOCK 2 BK 33364 17 SLAB DOOR ASSEMBLY 2 15147 18...

Страница 8: ...Pedigo P 2040 S Blanket Warming Cabinet Operation and Care Manual 8 P 2040 S Bonnet Service View TT 33563 w E3037TT SN 33540 BK 33546 BK 33364 NOT SHOWN TN 3972 FU 33332 E3006FU TT 33563 w E3037TT...

Страница 9: ...ation and Care Manual 9 P 2040 S Cabinet Service View E2088PE GS 23794 E4015 E1006 SN 33540 SW 3962 TT 33563 w E3037TT 4460 1844 1897 CASINGS 15147 HD 24171 4265 4572 CS 2026 FRONT CASTERS SWIVEL w BR...

Страница 10: ...Pedigo P 2040 S Blanket Warming Cabinet Operation and Care Manual 10...

Страница 11: ...hever occurs first The parts warranty on all other parts remains in effect three 3 years from installation or thirty nine 39 months from the shipping date whichever occurs first This warranty does not...

Отзывы: