IEEE488.2 Common Commands
26
CONFigure[:SCALar]:CAPacitance
Syntax
CONFigure[:SCALar]:CAPacitance [{<range>|MINimum|MAXimum|DEF|AUTO}]
Description
Restore all the measurement and trigger parameters to default, process capacitance
measurement. Then set the scale.
Parameter
Name
Type
Range
<range> Discret
e
2E-9(2nF), 20E-9(20nF), 200E-9(200nF), 2E-6(2uF), 20E-6(20uF), 200E-6(200uF),
10E-3(10mF)
CONFigure[:SCALar]:TEMPerature:RTD
Syntax
CONFigure[:SCALar]:TEMPerature:RTD [{<RTD Type>}]
Description
Restore all the measurement and trigger parameters to default, process temperature
measurement. Then set the RTD type.
Parameter
Name
Type
Range
<RTD Type >
Discret
e
KITS90, NITS90, EITS90, JITS90, TITS90, SITS90, RITS90, BITS90, W5_26,
W3_25, PT100, PT10, Cu100, Cu50