* Do not place the equipment face-down on any table or work bench to prevent damaging the controls at the
front.
* Opening the equipment and service – and repair work must only be performed by qualified service
personnel
*
-Measuring instruments don’t belong to children hands.-
Cleaning the cabinet
Clean only with a damp, soft cloth and a commercially available mild household cleanser. Ensure that no
water gets inside the equipment to prevent possible shorts and damage to the equipment.
1.1. Input limit
Function
Terminal
Max. input
limit
V DC
V/
Ω
+ COM
1000 V DC
V AC
V/
Ω
+ COM
750 V AC
Ω
V/
Ω
+ COM
250 V
AC/DC
mA-Bereich
AC/DC
MA + COM
220
mA
AC/DC
20 A-Bereich
20 A + COM
20 A
AC/DC
♫
V/
Ω
+ COM
250 V
AC/DC
Frequency
V/
Ω
+ COM
250 V
AC/DC
Logic
V/
Ω
+ COM
250 V
AC/DC
1.2. Safety Symbols
The maximum current that you can measure at this terminal is 20 A DC/AC. This terminal is
fuse protected by FF 15 A/250 V fuse. When using this range with high current, keep the duty
cycle to 30 Seconds on load and 15 Minutes off load.
The maximum current that you can measure with this terminal is 220 mA DC/AC. This terminal
is fuse protected by 0,8 A/250 V fuse.
To avoid electric shock or instrument damage, do not connect the common input COM
terminal to any source of more than 500 V with respect to earth ground.
The maximum voltage this meter can measure is 1000 V DC or 750 V AC.
Be exceptionally careful when measuring high voltages. Do not touch the terminals or test lead
ends!
Refer to the complete operating instructions.
Indicates protection class II, Double insulation.
CAT II
Overvoltage category II.
-16-