2.2. Electrical Specifications
DC Volts
Model
Range
Resolution
Accuracy
P 3355
400 mV
100 µV
±
0,5% rdg. + 2 dgt
4 V
1 mV
±
1,2% r dg. + 2 dgt
40 V
10 mV
400 V
100 mV
1000 V
1 V
±
1,5% rdg. + 2 dgt
P 3360
400 mV
10 µV
4 V
100 µV
±
0,06% rdg. + 2 dgt
40 V
1 mV
400 V
10 mV
1000 V
100 mV
±
0,1% rdg. + 5 dgt
Overload protection
1000V AC/DC
rms
Input Impedance:
(P 3355) 7,8 M
Ω
/ (P 3360) 10 M
Ω
AC Volts
Model
Range
Resolution
Accuracy
P 3355
400 mV
100 µV
±
1,0% rdg. + 5 dgt
4 V
1 mV
±
1,0% rdg. + 3 dgt
40 V
10 mV
400 V
100 mV
1000 V
1 V
P 3360
400 mV
100 µV
±
1,0% rdg. + 5 dgt
4 V
1 mV
±
1,0% v.M. + 3 dgt
40 V
10 mV
400 V
100 mV
1000 V
1 V
Overload protection
1000V AC/DC
rms
Input Impedance:
(P 3355) 7,8 M
Ω
/ (P 3360) 3 M
Ω
AC Response:
(P 3355) 50 to 400Hz / (P 3360) 50 to
1000Hz
-40-