Document No.:
in the calibration confirmation interface prompt box to start Auto Calibrate. Auto
Calibrate progress interface pops up at the same time. Press cancel key
on the
Auto Calibrate confirmation interface. Return to the Auto Calibrate interface.
Step III: Auto Calibrate progress interface (Figure 3). The Auto Calibrate progress prompt is
displayed on the interface, and the calibration result interface appears after the Auto
Calibrate is completed. Press cancel key
on the Auto Calibrate progress interface.
Return to the Auto Calibrate interface.
Step IV: The calibration result interface (Figure 4) appears after the Auto Calibrate is
completed. Tick √ in the prompt box for success and tick × in the prompt box for failure.
Press cancel key
on the Auto Calibrate result interface. Return to the Auto
Calibrate interface.
Note: The Auto Calibrate must be carried out at full charge.
Figure 1 M4-HD system information view interface
A: System information view item window prompt
B: Select the system information view item from the setting list
A
B