Chapter VII Maintenance and Care
To ensure the normal operation of the instrument, the instrument parts, accessories
and probe shall be cleaned and maintained regularly. The material used for cleaning is
neutral detergent.
7.1 Cleaning and maintenance of the system
7.1.1 Cleaning and disinfection steps
(1) Turn off the system.
(2) Use mild, nearly neutral detergent, ethanol (75%) or isopropanol (70%) to clean
equipment (including keyboard).
(3) Control the time of wiping according to the instructions for use of detergent and the
interval of wiping shall meet clinical requirements.
(4) Dry the surface of the system or dry it with cleaning cloth according to the method
described on the detergent label.
(5)
Clean fingerprints or other contaminants on the display screen and wipe the
display screen with soft and damp cleaning cloth dipped with neutral detergent to ensure
that the display screen will not be scratched.
7.1.2 Maintenance
(1) The instrument shall be operated in the environment specified in "
1.5
".
(2) The instrument shall not be turned on or turned off frequently. After being turned off,
it shall be turned on after 5 minutes.
(3) When the instrument is not in use for a long time, it shall be packed well according
to the factory packaging standard and stored in accordance with the storage environment
requirements specified in "
8.2
".