C-7100 Series User Manual PEAK INSTRUMENTS INC.
20
VII. Time scan (Kinetic)
1. Function description
Scan time (kinetic) function is a fixed time interval to the trend in absorbance or transmittance of the test current,
and displayed on the map. Select time scanning (kinetic) menu option press [ENTER].
2. Time scanning parameters set
Set the time to scan the scanning parameters: time interval, test time, measurement mode, the display displays the
upper and lower limits
1) Set test time
Test time is the total time of the entire test.
2) Set the measurement mode
The measurement mode is absorbance, transmission or energy. Choose different measurement mode, you need to
re-set the display to display the upper and lower limits.
3) Set upper and lower limit
The different measurement modes, upper and lower display is not the same.
4) Set the time interval
Scan scanning interval setting time, 0.5 seconds minimum, 1 minute maximum.
5) Select to make data smoothing
The function for data smoothing is to reduce the irregular fluctuation caused by external environment during test.