Zero Air Gas Generator 

Instructions for use Manual 

30

th

 March 2005 

 

ZA015 Issue 3

 

 

Introduction

 

 

The Peak Scientific Instruments range of Zero Air Gas Generators is designed to produce a constant flow of 
Zero Grade Air with a Hydrocarbon content (as Methane) of less than 0.1 ppm. 

 

Unpacking and Installation 

 

Although  Peak  Scientific  take  every  precaution  with  safe  transit  and  packaging,  it  is  advisable  to  fully 
inspect the unit for any sign of transit damage.

  

 
ANY  DAMAGE  SHOULD  BE  REPORTED  IMMEDIATELY  TO  THE  CARRIER  AND 
PEAK  SCIENTIFIC  OR  THE  DISTRIBUTOR  FROM  WHERE  THE  UNIT  WAS 
PURCHASED.

 

 

After unpacking and a visual inspection, the unit should be placed in a ventilated area away from direct 
sunlight. Care should be taken  not to obstruct the ventilation holes  on the sides of  the unit nor, on the 
ZA180, the fan outlet on the top. 
 
The generator should be placed on a steady and level base. Alternatively, the unit may be wall mounted 
using the slots on the rear of the cabinet. These are on 100mm centres. 
 

Electrical Connection

 

 

Important Electrical Notice 

 

This  unit  is  classified  as  SAFETY  CLASS  1  equipment.  THIS  UNIT  MUST  BE  EARTHED.  Before 
connecting  the  unit  to  the  mains  supply,  please  check  the  information  on  the  serial  plate.  The 
mains supply must be of the stated AC voltage and frequency.

 

 
EARTH/GROUND (E): - 

 

Green & Yellow 

or 

Green 

LIVE (L): - 

 

 

 

Brown 

 

 

Black 

Neutral (N): -  

 

 

Blue   

 

 

White

 

 

Fuse

 

The generator protection fuse in the pull out drawer of the mains inlet IEC euro connector located 
on the bottom right hand side of the cabinet adjacent to the off/on switch. The fuse is rated at 6.3 
AMP. 
 

Connect the generator to a single-phase supply using the power cord provided. 
 

Содержание ZA015

Страница 1: ...as Generator ZA015 ZA180 Peak Scientific Instruments Ltd Fountain Crescent Inchinnan Business Park Inchinnan Renfrewshire PA4 9RE T Europe 44 0 141 812 8100 T USA 1 866 647 1649 sales T USA 1 866 732...

Страница 2: ...Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 2 Document Change History Issue No Changed Initials Date 1 Document Created SGM 01 09 98 2 USA Technical support number updated FAD...

Страница 3: ...Issue 3 3 Contents Page No Document History 2 Warranty Statement 5 Introduction 6 Unpacking Installation 6 Electrical Connection 6 Air Connection 7 Principle of Operation 8 Commissioning 9 Routine Mai...

Страница 4: ...persons should operate or service this equipment Warning To avoid risk of electrical shock personal injury or death disconnect power before removing any cover of this equipment Caution To avoid risk o...

Страница 5: ...he Goods and or any installation and any relevant packaging as and when reasonably required by the Company 5 Subject to the Company being notified of any defect as is referred to in sub clause 2 hereo...

Страница 6: ...ould be taken not to obstruct the ventilation holes on the sides of the unit nor on the ZA180 the fan outlet on the top The generator should be placed on a steady and level base Alternatively the unit...

Страница 7: ...isoned if it comes into contact with any halogenated hydrocarbons silicone sprays silicone greases phosphorous compounds lead components high sulphur vapours or other catalyst poisons The air supply s...

Страница 8: ...isture and particles down to 0 1 micron The air is then passed to the Zero Air catalytic combustion chamber This works on the principle of catalytic oxidation where hydrocarbons from the incoming comp...

Страница 9: ...r a period of time around 40 minutes the Catalytic Chamber should have reached the required temperature When this happens the relay on the Temperature Controller will click and the red LED will go out...

Страница 10: ...increased levels of Hydrocarbons being detected on the baseline Reference to the following Fault Finder Chart will identify the source of the problem High CH4 Cat Chamber Hot Cold Control Board LED LE...

Страница 11: ...5 3 5 7 0 18 0 Heater Watts 250 250 250 450 Fan Watts N A N A N A 9 0 Electrical Requirements 230V 50Hz 110V 60Hz 1 2A 2 3A 1 2A 2 3A 1 2A 2 3A 2 0A 4 5A Dimensions H x W x D cm Inches 51x25x16 20x10...

Страница 12: ...2005 ZA015 Issue 3 12 10 System Diagrams Heater Rod Thermocouple Outlet ZAxxx Pneumatic Diagram Inlet Filter Regulator Cat Chamber Cooling Coil Thermocouple Heater Rod Out N L Temperature Controller 4...

Страница 13: ...PEAK Scientific Instruments Ltd Zero Air Gas Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 13 Maintenance Log Model Serial number Work Done Remarks Date Name...

Страница 14: ...PEAK Scientific Instruments Ltd Zero Air Gas Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 14 Notes...

Отзывы: