TOC 1500 

User Manual 

Page 11  

2014 © Peak Scientific - Rev. 3A - 18.11.14 

Installation 

Generator Environment 

 

The TOC 1500 Gas Generator is designed for indoor use only. 
The unit can be free standing, in which case it must be placed 
on a steady and level surface. 

 
The minimum and maximum ambient operating temperatures are 

5

o

C and 

25

o

C. 

Keep the unit in a well ventilated site away from direct sunlight. 

Voltage and Power 

The TOC 1500 Gas Generator are fitted with an internal switch 
mode    power supply that will accept a power inlet from 
100VAC - 240VAC, 50/60HZ. The current consumption is 80 
mA.  

 

General Dimensions

 

 

 

472.4 

660 

160 

100 

570 

TOC 1500 

Содержание TOC 1500

Страница 1: ...TOC 1500 User Manual 2014 Peak Scientific Rev 3A 18 11 14...

Страница 2: ...rmity 7 Environmental Declaration 8 Technical Specification 9 Environment 9 Outlet Gas 9 Electrical Requirements 9 General 9 Unpacking 9 Installation 11 Generator Environment 11 Voltage and Power 11 G...

Страница 3: ...e or as a reference document where you can skip to the relevant information Users of a hard copy version can refer to the contents page to find the relevant information Users of the soft copy version...

Страница 4: ...re that your Dryer is operated according to our recommended guidelines which will prepare you for long and trouble free life Attachable to any compressed air source Peak Scientific s TOC 1500 Gas Gene...

Страница 5: ...ithout prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and or servicing incorrect repair and or servicing use of wrong materials and...

Страница 6: ...egulations The attention of UK users is drawn to the Health and Safety at Work Act 1974 and the Institute of Electrical Engineers regulations A WARNING notice denotes a hazard It calls attention to an...

Страница 7: ...TOC 1500 User Manual Page 7 2014 Peak Scientific Rev 3A 18 11 14 Declaration of Conformity Unit is currently under accreditation...

Страница 8: ...TOC 1500 User Manual Page 8 2014 Peak Scientific Rev 3A 18 11 14 Environmental Declaration Unit is currently under accreditation...

Страница 9: ...emperature for a minimum of 3 hours before operation Outlet Gas Output Flow Rate 105 L min Pressure 5 psi Below Inlet Pressure Pressure Dewpoint 70 C 100 F Particles 0 01 um Startup Time 16 Hours Elec...

Страница 10: ...structions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvre the Generator onto the floor Please save the product packaging for sto...

Страница 11: ...laced on a steady and level surface The minimum and maximum ambient operating temperatures are 5 o C and 25 o C Keep the unit in a well ventilated site away from direct sunlight Voltage and Power The...

Страница 12: ...nput port of the unit on the left hand side of the unit Attach a suitable air supply to the inlet port of the filter Note that the inlet and outlet air connections are BSP Female Attach the outlet of...

Страница 13: ...pecification Ensure that the power to the Air Dryer is turned on before starting the air supply Gradually introduce air pressure to the unit until the recommended pressure has been reached Allow unit...

Страница 14: ...EMAIL ADDRESS CITY TOWN GENERATOR SERIAL NUMBER POSTCODE COUNTRY MODEL TYPE TELEPHONE INSTALLATION DATE DD MM YYYY Important Please Note You have 1 month to register your Peak Scientific product from...

Страница 15: ...let air to the Dryer passes through a pre filtration process where it undergoes particulate removal 0 01 micron and bulk moisture removal by a coalescing element The air is then passed to the sieve co...

Страница 16: ...with the generator in its safely isolated condition Safely Isolated Condition The unit is in a Safely Isolated Condition when it is disconnected from its application fully de pressurised and isolated...

Страница 17: ...ponents separately Purchase Interval Component Part No Qty 12 months TOC 1500 Annual Service Kit 08 8222 1 Service Plans Peak Scientific offer two service plans The Complete Service Plan specifically...

Страница 18: ...1 Lane 2277 Zuchongzhi Road Pudong New Area Shanghai 201203 China Peak Scientific Mexico Solon 352 Col Los Morales Polanco 11530 Mexico D F Mexico Peak Scientific Australia PO Box 65 Belmont Victoria...

Отзывы: