12.
11.
45
minutes
Leave the generator
switched on
for a period of
45 minutes
to purge the system.
Remove the front panel
and check gauges have risen to the factory set pressures
stated in the
User Manual. The gauges are located in the bottom left hand corner
of the G SCI 1. For G SCI 2 models, the gauges are located behind the fascia panel
which can be removed by pulling towards you.
Replace
any front panels
before continuing.
G SCI 2
G SCI 1