Troubleshooting

Problem

Possible Solution

The Generator will not switch on and the 

power switch does not illuminate.

•  Ensure power cord is plugged  

into the Generator and that  

the power socket is turned on.

•  Check the fuse in the power cord plug.
•  Check for any disconnected terminals at 

power socket connected at inside rear of 

unit.

•  Check for loose connections at inside of 

power switch located on front of Generator.

•  Check for 230VAC  going into DC power 

supply

•  Check for 24VDC  coming from power 

supply

•  Check circuit breaker is ON – switched up
•  Contact Peak Scientific Technical Help. 

The Generator will not switch on but the 

power switch is illuminated.

•  Disconnect power cord from  

the rear of the Generator.  

Open the left hand panel 

•  Check that circuit breakers  

are turned on (switch in the up position). 

Reconnect power cord.

•  Check that each compressor has its 

associated electrical plug connected.

•  Check compressor connectors on PCB for 

loose wires.

•  Check electrical panel for loose wires.
•  Check pressure switch for loose wires.
•  Check pressure switch LED is lit.
•  Contact Peak Scientific Technical Help.

Compressor is running but pressure is not 

building in system.

•  Check drain outlet at rear of Generator is 

not passing air.

•  Open the left hand panel and check if tank 

pressure gauge is rising, thus pressure 

being built in receiver tanks.

•  Check for leaks in system.
•  Contact Peak Scientific Technical Help.

Содержание MS Bench SCI 1

Страница 1: ...MS Bench G SCI 1 Service Manual Rev 3 RSID 3025 EN 20 11 20 MS Bench G SCI 1 Service Manual...

Страница 2: ...tion 8 Service Requirements 9 Service Schedule 9 Product Overview 10 PCB Electrical Panel Layout 11 Compressor Inlet Filter Element Replacement 12 Twin Filter Element Replacement 13 RAC Filter Element...

Страница 3: ...end users and has been written as a reference document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review each of the f...

Страница 4: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 5: ...g the generator are shown below Failure to place the generator in a safely isolated condition when instructed to do so may lead to personal injuryor injury to others and even death Isolating the gener...

Страница 6: ...ould be plugged into the IEC power inlet located on the rear panel of the generator This unit is classified as SAFETY CLASS 1 THIS UNIT MUST BE EARTHED The system builds pressure correctly With the ou...

Страница 7: ...ource Maximum Flow 26 L min 6 89 bar 0 918 cfm 100 psi Exhaust Maximum Flow 25 L min 4 13 bar 0 882 cfm 60 psi Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 3 x BSPP Drain Outlet...

Страница 8: ...d The air division is split once more between the desired outputs SOURCE and EXHAUST Finally all output product is passed through an RAC filter prior to delivery to the application Once the designated...

Страница 9: ...ntenance Kit www peakscientific com ordering 4 Years MS Bench G SCI 1 Year 4 Service Kit www peakscientific com ordering Please quote your generator serial number when ordering your Maintenance Kit to...

Страница 10: ...ew PCB Membrane RAC Filters Twin Filters Compressors Valves Power Inlet Switch Regulators Dryer Filter Capacitors Valves Reservoir Tanks Air Inlet Filter Electrical Panel Outlet Panel Curtain Source E...

Страница 11: ...f electrics shock accessing the electrical panel without safely isolation the generator should only be carried out by trained service personnel WARNING Power Supply Pressure Switches Circuit Breaker P...

Страница 12: ...n 1 Release the filter bowl by turning it anticlockwise 2 Remove the bowl by pulling it towards you 3 Remove the paper and foam filters and clean the inside of the filter bowl before fitting replaceme...

Страница 13: ...y Isolated Condition 1 Remove the 6mm drain tubing for push fit stud situated on the underside of both of the bowls 2 Remove the filter bowl by turning anti clockwise 3 Remove the filter elements by p...

Страница 14: ...Safely Isolated Condition 1 Remove the 6mm drain tubing for push fit stud situated on the underside of the bowl 2 Remove the filter bowl by turning anti clockwise 3 Remove the filter element by pullin...

Страница 15: ...y Isolated Condition 1 Remove the 6mm drain tubing for push fit stud situated on the underside of both of the bowls 2 Remove the filter bowls by turning anti clockwise 3 Remove the filter elements by...

Страница 16: ...4 Remove the two screws located in the top corners of the compressor cover 5 Disconnect the drain and offloader tubes from the valves Unclip the panel by pulling the panel from the top towards you 6...

Страница 17: ...essure switch is for indication only calibration is not required Nitrogen Pressure Switch Settings Output Set Point SP Hysteresis HY 1 Refer to QD09 Refer to QD09 Green Output OUT 1 Red Output OUT 2 B...

Страница 18: ...N DOUBT ASK PEAK SCIENTIFIC INSTRUMENTS Ltd 2013 PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRITTEN CONSEN...

Страница 19: ...DATE ELECTRICAL SCHEMATIC FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WIT...

Страница 20: ...HEMATIC FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRITTEN CO...

Страница 21: ...L SCHEMATIC FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRITTE...

Страница 22: ...CAL SCHEMATIC FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRIT...

Страница 23: ...SCHEMATIC FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRITTEN...

Страница 24: ...FOR PROJECT LEAD ENGINEER DRAWING No DCL REV PEAK SCIENTIFIC OWN THE COPYRIGHT FOR THIS DRAWING NO INFORMATION CONTAINED IN THIS DRAWING MAY BE CONVEYED TO A THIRD PARTY WITHOUT THE WRITTEN CONSENT O...

Страница 25: ...POINT NO YES IS THIS THE FIRST TIME AT PRESSURE YES NO STORE DATE IN MEMORY IS FORCED ON FLAG SET FOR AVAILABLE COMPRESSOR NO YES START COMPRESSOR SHUT DOWN SEQUENCE FOR AVAILABLE COMPRESSOR START NOR...

Страница 26: ...OFFLOADER START NORMAL OPERATION CLOSE OFFLOADER START COMPRESSOR SHUT DOWN SEQUENCE START START UP SEQUENCE START 1 MIN AT PRESSURE SEQUENCE ARE 1 MIN PRESSURE FLAG FORCED ON FLAG FORCED OFF FLAG ALL...

Страница 27: ...NCE OPEN OFFLOADER CLOSE OFFLOADER OPEN DRAIN CLOSE DRAIN HALT DRAIN FAILSAFE SEQUENCE END COMPRESSOR START UP SEQUENCE OPEN OFFLOADER 1 SEC DELAY START COMPRESSOR START FORCED ON SEQUENCE 1 SEC DELAY...

Страница 28: ...D FORCED OFF SEQUENCE START 3 MIN TIMER HAS TIMER EXPIRED NO YES SET FORCED OFF FLAG END 1 MIN AT PRESSURE SEQUENCE START 1 MIN TIMER HAS TIMER EXPIRED NO YES SET 1 MIN AT PRESSURE FLAG END 5 MIN AT P...

Страница 29: ...Page 29 DRAIN FAILSAFE SEQUENCE START 15 MIN TIMER HAS TIMER EXPIRED NO YES OPEN DRAIN VALVES FOR 1 SEC CLOSE DRAIN VALVES START DRAIN FAILSAFE SEQUENCE...

Страница 30: ...e cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cle...

Страница 31: ...or will not switch on but the power switch is illuminated Disconnect power cord from the rear of the Generator Open the left hand panel Check that circuit breakers are turned on switch in the up posit...

Страница 32: ...aks between Generator output and Mass Spectrometer Contact Peak Scientific Technical Help Yellow Service Indicator is on constantly Compressor is overdue for service Replace compressor refer to Compre...

Страница 33: ...Page 33...

Страница 34: ...Page 34...

Страница 35: ...Page 35...

Страница 36: ...d Service Engineers located in over 20 countries across every continent around the world This allows us to provide an industry leading rapid response service to our customers With Peak Protected your...

Отзывы: