Service Requirements 

Service Schedule  

Purchase Interval 

Component 

Part No. 

Qty. 

12 months 

Detector Annual Service Kit 

08-9003 

Table 1: Annual Service Kit 

Trouble Shooting 

Fault Code 

Possible Solution 

M10 Over Range – A large 
concentration has been detected. 
The Detector requires an 
independent confirmation that 
the gas hazard has gone. 

 

Supply known clean air to the Detector, and clear this fault. 

 

Contact Peak Service Provider. 

M12 Cartridge Expires Soon – 
Cartridge is old and will expire 
soon. 

 

Replace the cartridge with a new cartridge. 

 

Contact Peak Service Provider. 

M13 Flow Error – Detector is no 
longer able to regulate flow (high 
or low flow condition). 

 

Check filters and pump. 

 

Check pressure at inlet and outlet and assure they are within 
specification. 

 

Contact Peak Service Provider. 

M14 Interferent Present – An 
interferent is degrading the ability 
of the Detector to detect gas. 

 

Check application. 

 

Contect Peak Service Provider. 

M15 Temperature Near Limit – 
Temperature within 2 Celsius of 
limit. 

 

Check installation environment. 

 

Contact Peak Service Provider. 

M16 Baseline Fault – Sensor 
baseline has drifted. 

 

Check for background gas concentration, temperature or 
humidity fluctuations.  

 

Perform zero calibration.  

 

Replace cartridge. 

 

Contact Peak Service Provider. 

M17 Inhibit Timeout – Transmitter 
has been in inhibit mode too long 

 

Resume monitoring or increase timeout value. 

 

Contact Peak Service Provider. 

F40 Sensor Overdosed – Sensor 
has been exposed to high gas 
concentrations for long periods. 

 

Replace Cartridge. 

 

Contact Peak Service Provider. 

F43 Cartridge expired – Cartridge 
is too old. 

 

Replace Cartridge. 

 

Contact Peak Service Provider. 

F44 Cell Failure – Cartridge has 
failed reflex check. 

 

Replace Cartridge. 

 

Contact Peak Service Provider. 

F45 Stabilisation Timeout – 
Cartridge has failed to stabilise. 

 

If temperature or humidity shocks exist, precondition the 
cartridge. 

 

Check background gas concentration. 

 

Replace Cartridge. 

Содержание Hydrogen Detector

Страница 1: ...Hydrogen Detector User Manual 2021 Peak Scientific Rev 5 12 10 2021...

Страница 2: ...pecification 9 Environment 9 Detector Inlets 9 Default Configuration 9 Electrical Requirements 9 General 9 Unpacking 10 Installation 11 Detector Environment 11 General Dimensions 12 Rear Connections 1...

Страница 3: ...llation and usage or as a reference document where you can skip to the relevant information Users of a hard copy version can refer to the contents page to find the relevant information Users of the so...

Страница 4: ...ven to a sensor cartridge that is located inside the detector s chassis Our Hydrogen Detector is wall mountable and displays gas concentration alarm fault and status information via its backlit LCD an...

Страница 5: ...to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and or servicing incorrect repair and or servicing use of wrong materials and or incorrect spa...

Страница 6: ...k Act 1974 and the Institute of Electrical Engineers regulations If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment maybe impaired A WARNIN...

Страница 7: ...2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for Measurement Co...

Страница 8: ...ety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibility Regulations 2016 S...

Страница 9: ...ance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE marking...

Страница 10: ...lue TLV 100 Lower Explosive Limit LEL 25 Volume v v Lowest Alarm Level LAL Typically TLV 10 LEL 5 v v Lower Detectable Level LDL Typically 0 4 TLV 9 LEL 0 v v The LDL is the minimum level that is reli...

Страница 11: ...of rough handling prior to un packing Any damage should be reported immediately to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased Please save the product packagi...

Страница 12: ...could adversely affect its performance Please refer to the drawing below for the general dimensions of the unit Maximum Ambient Conditions 40 C dry bulb 90 RH Max Non Condensing Location The detector...

Страница 13: ...H Detector User Manual Page 12 2013 Peak Scientific Rev 4 03 09 1 General Dimensions The Detector must always be placed on a level surface Failure to do so will affect the performance of the Detector...

Страница 14: ...HK Exhaust Tubing 1 5m 6 OD x 3 16 THK Input Tubing 1 5m 7 OD Stainless Steel Tubing x 1m 8 Bulkhead Reducer Fitting x 1 9 Tygon Filter Tubing x 1m 10 Hex Key x 1 11 Hose Clips x 2 12 Wall Mounting Br...

Страница 15: ...late directly above the rear hole on the GC oven 3 Make a hole through the insulation material through to the oven The stainless steel tubing from the detector can now be pushed through the hole and s...

Страница 16: ...alarm if dangerous levels of Hydrogen are detected The generator will then shut down Tubing Lengths The length of the tubing which will be connected to the Input Port of the Detector is important and...

Страница 17: ...cation between the Precision 500 and the detector unit The detector will take samples continuously and both the detector and the Precision 500 will alarm if dangerous levels of hydrogen are detected t...

Страница 18: ...in place by two locking tabs Some cartridges are shipped with a protective cap to shield them from contaminants during shipping This cap must be removed before inserting the cartridge into the Detecto...

Страница 19: ...he display hit the on the Detector front panel 2 When the reboot completes then press and hold the X to clear any latched fault s 3 Confirm that the green LED is flashing 4 Confirm that the yellow and...

Страница 20: ...messages WAIT and LOAD as it checks for the cartridge data typically less than 180 seconds 4 When complete the detector will enter normal monitoring mode indicated by the icon on the display cycling...

Страница 21: ...EMAIL ADDRESS CITY TOWN DETECTOR SERIAL NUMBER POSTCODE COUNTRY MODEL TYPE TELEPHONE INSTALLATION DATE DD MM YYYY Important Please Note You have 1 month to register your Peak Scientific product from...

Страница 22: ...rent is degrading the ability of the Detector to detect gas Check application Contect Peak Service Provider M15 Temperature Near Limit Temperature within 2 Celsius of limit Check installation environm...

Страница 23: ...Cartridge Replace Cartridge Contact Peak Service Provider F49 Cartridge Wrong Type Cartridge type found to be incorrect after boot up Replace Cartridge or press to accept if correct Contact Peak Serv...

Страница 24: ...1 Lane 2277 Zuchongzhi Road Pudong New Area Shanghai 201203 China Peak Scientific Mexico Solon 352 Col Los Morales Polanco 11530 Mexico D F Mexico Peak Scientific Australia PO Box 65 Belmont Victoria...

Отзывы: