Service Indication

The generator will notify the user of the service interval for the internal compressors. 

The generator has the following Service Indication Stages:-

Stage 1 

12 months after installation, the service indicator will show on the display along with a 

warning message.
This is to make the user aware that a service of the generator is due and should be 

planned at the earliest convenience. The generator will continue to operate as normal 

with the service indicator on. If the warning message has been acknowledged, the icon 

will still be displayed in the corner of the HMI; pressing the icon will show the message 

again.

Stage 2

 

If the service is not completed the generator will continue to run. After 2 weeks, the 

service overdue indicator will show on the HMI along with a warning message.
This is to make the user aware that the service of the generator is now overdue and 

must be completed immediately to ensure the continuous trouble free operation of 

the generator. If the warning message has been acknowledged, the icon will still be 

displayed in the corner of the HMI; pressing the icon will show the message again.

Service Indication Reset

Once the service has been completed the Service Indication can be reset through the 

service interface. This will be performed by the Peak Service Engineer or trained service 

representative that completes the service operation.

Содержание Genius XE 70

Страница 1: ...Genius XE All Models User Manual Copyright Peak Scientific Instruments Ltd 2020 GeniusXE User Manual Rev 9 RSID 2299 EN 18 02 2020...

Страница 2: ...ranty your generator must be registered to the end user not a reseller or distributor and must have a paid annual preventative maintenance arranged within 12 months of the installation date carried ou...

Страница 3: ...13 Generator Overview 14 Drain Connection 16 Electrical Connection 17 Start Up Sequence 18 Product Registration 19 Connecting to the application 21 Normal Operation 23 Eco Mode 23 Service Screens 24...

Страница 4: ...Update Liam Couttie 16 12 2019 9 Technical Specification Update Liam Couttie 18 02 2020 How to use this Manual This manual is intended for end users and has been written as a reference document where...

Страница 5: ...upply to the generator must comply with ISO 8573 1 2010 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partne...

Страница 6: ...conditions outlined in this warranty state ment in any respect or to create any additional obligations or liabilities for any party involved 11 This warranty statement supersedes any and all prior war...

Страница 7: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 8: ...nt for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for measurement control and laboratory use FCC47 CFR Part 15 Class...

Страница 9: ...h confirm that as a manufacturer of articles we conform to the European Parliament and Council Directives 2011 65 EU for the original 6 restricted substances listed in the Restriction of Hazardous Sub...

Страница 10: ...0V 10 230V 10 Frequency 120V 60Hz 230V 50 60Hz 230V 50 60Hz Current 12A 120V 8A 230V 12A 230V Input Connection C20 Plug Power Cord C19 Socket Circuit Breakers 230V 6A MCB 2A MCB 120V 10A MCB 2A MCB Po...

Страница 11: ...the Peak Partner from where the unit was purchased Follow the unpacking instruction supplied with the generator It will require two people to remove the unit from the shipping crate and to manoeuvre...

Страница 12: ...1 8 US 120V Mains Power Cable x 1 9 Genius XE Installation Guide x 1 Genius XE 35 XE 70 230V 1 1 4 x 1 4 Compression Fitting x 1 2 1 4 x 6mm Push Fit Fitting x 2 3 Flow Control Silencer x 1 4 1 4 PTF...

Страница 13: ...et could adversely affect its performance Consideration should also be given to the air flow around the unit An air gap of 75mm 3 should be maintained down both sides and 100mm 4 at the rear of the un...

Страница 14: ...mensions Genius XE 35 Genius XE 70 The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 570 mm 22 4 710 mm 27 9 650 mm 25 6...

Страница 15: ...ear Connections Ensure outlets are connected to correct applications Connections should only be carried out by trained personnel Unit Controls Touch Screen Nitrogen Port Drain Port Power Socket WARNIN...

Страница 16: ...ain connection or container It should be noted that the generator can expel a considerable amount of water dependant on ambient humidity If a container is used it should be emptied at regular interval...

Страница 17: ...te The mains supply must be of the stated AC voltage and frequency EARTH GROUND E Green Yellow or Green LIVE L Brown or Black Neutral N Blue or White Electrical requirements are 120VAC nominal 5 or 23...

Страница 18: ...and switching it on for the purge run it is necessary to fit the silencer to the nitrogen output port Once this is done the generator can be connected to the mains and switched on The Genius XE will t...

Страница 19: ...also be requested by phoning the Peak helpdesk This code is only required on the initial start up of the unit 1 Press the I ve got my pin button on the screen on the front of the generator 2 Enter you...

Страница 20: ...Page 20 4 Upon completion of the initialisation phase the screen will display the below home screen GENIUS XE 35 S N XXXXXXXXXXXX Nitrogen XXXX psi...

Страница 21: ...lease ensure that it is thoroughly checked for being leak tight Even the slightest leak in the gas supply between the generator and the application can lead to a drop in Nitrogen purity or insufficien...

Страница 22: ...116psi The output pressure can be adjusted using the pressure regulator at the rear of the generator and viewed on the adjacent gauge To ensure a smooth pressure profile it is not recommended to incre...

Страница 23: ...5 nitrogen at the maximum rated output flow Eco Mode To reduce energy consumption and minimise component wear Genius XE generators feature a smart Eco Mode which ensures gas is produced as is necessar...

Страница 24: ...XXXXXXXXXXX Nitrogen XXXX psi Service Due Home Screen Warning This is the main home screen when a service is due displaying a yellow service warning The screen will alternate between the home screen a...

Страница 25: ...ng as shown next GENIUS XE 35 S N XXXXXXXXXXXX Nitrogen XXXX psi REF CODE XXXX REF CODE XXXXXXXXXX OR VISIT WWW PEAKSCIENTIFIC COM SERIAL XXXXXXXXXXXX Confirm Minor Error Warning Screen This is the fu...

Страница 26: ...100 l min Temp Hum Temp Hum 10 xx xx xx xx xx xx xx xx xx xx xx xx xx xx 0 10 20 30 40 C 50 60 70 80 90 0 10 20 30 40 50 60 70 80 90 100 Temperature Graph 2 results plotted Temperature Humidity Y axi...

Страница 27: ...ct Pressure Unit bar atm kPa Select Flow Unit l min l hour m3 min m3 hour Select Flow Unit Screen Allows the user to select a unit of Flow Select Pressure Unit Screen Allows the user to select a unit...

Страница 28: ...l Component Visit Qty 12 Months Genius XE 35 120 230V Annual Maintenance Kit www peakscientific com ordering 1 Genius XE 70 230V Annual Maintenance Kit 1 48 Months Genius XE 35 120V 48 Month Kit 1 Gen...

Страница 29: ...essing the icon will show the message again Stage 2 If the service is not completed the generator will continue to run After 2 weeks the service overdue indicator will show on the HMI along with a war...

Страница 30: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 31: ...he cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cl...

Страница 32: ...d contact your service provider Minor Alarm Message There is an issue with the generator but it will continue to supply gas at the required pressure flow and purity Major Alarm Message There is a prob...

Страница 33: ...tact your service provider Service indicator on the screen is active The generator is due for service Contact your service provider Refer to Service Indication section of this manual for further infor...

Страница 34: ...Page 34 Notes...

Страница 35: ...Page 35 Notes...

Страница 36: ...ed in over 20 countries across every continent around the world This allows us to provide an industry leading rapid response service to our customers With Peak Protected your laboratory s productivity...

Отзывы: