Fountain Crescent
Inchinna Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W:
Part No. IG-Genius3040
Rev.
3
Page
Page 8 of 9
28.
29.
The
‘
CURTAIN
’ ‘SOURCE’ ‘EXHAUST’ &
‘NITROGEN’
pressure gauges on the front
of the generator will again gradually
increase from
0
to
80, 110, 60 & 100 psi
respectively,
as the pressure in the
internal storage tanks increases.
CONGRATULATIONS
Your
PEAK SCIENTIFIC
gas generator is now
fully installed, operational and ready to supply
gas on demand to your instrument.
General Notes
*
The generator will operate on voltages of /- 10%. This means it will
operate between 207V and 253V. If the generator is operated on a supply voltage of
less than 220V the internal compressors will work harder and the service life of the
compressors will be reduced. It is ok to operate the generator on a mains voltage
between 207V and 220V for a short period of time, however we would recommend
fitting the listed transformer at your earliest convenience. Mains voltages between
220V and 253V no transformer is required.
**
The 12m length of Teflon tube supplied in the fittings kit is to allow up to 3m of
tubing per output, for the generator to be installed adjacent to the instrument it is
supplying. If the generator is to be installed further away from the instrument then
attention needs to be paid to pressure drops that could occur. The maximum
distance away from the instrument the generator can be placed still using 6mm (4mm
internal diameter) or ¼” (3/16” internal diameter) tubing is 10m. For further
information relating to tubing lengths please refer to the user manual.
For all other technical specifications, operating instructions, service
requirements, contact details and trouble shooting, please refer to
the user manual contained on the CD supplied in the fittings kit.
Please keep this for future reference.
30.
It is very important to register your generator with PEAK SCIENTIFIC. This will initiate your
warranty entitlement. Please use the form on the next page to register your generator. You
will need the generators serial number which can be found on the serial label on the rear of
the generator.