6. 

       

 

7. 

          

 

Remove the LH side panel as before, 

taking care to disconnect the earth lead 

located on the inside of the door panel. 

Repeat steps 4 and 5 to remove the LH transit 

brackets located under both of the LH 

compressors. These will pull out in the same 

way. 

**NOTE** 

Do not discard the transit brackets or screws as these will be needed again if the                 

unit is to be transported from its current location. 

8. 

 

9. 

          

 

From the fittings kit, select the 6mm push-fit 

fitting and connect it to the port marked 

‘DRAIN’

 on the rear of the unit using a 16mm 

or 5/8” spanner. 

 

Connect the 2m Nylon tube to the drain 

line, ensure the tube is pushed fully in and 

gripped securely by the fitting. 

Fit the other end of the drain line to a 

suitable drain connection or container. The 

container must 

NOT

 have an airtight seal 

as water and air will be expelled 

periodically under a slight pressure.  

 

2m Nylon 
to Drain 

Содержание Genius 3040

Страница 1: ...iness Park Inchinnan PA4 9RE Scotland UK T 44 0 141 812 8100 F 44 0 141 812 8200 W www peakscientific com E info peakscientific com Part No IG Genius3040 Rev 3 Page Page 1 of 9 Installation Guide Geni...

Страница 2: ...the RH side panel Using the Hex Key provided in the fittings kit turn the locks 90o to the right An earth lead is connected to the inside of the panel This can be disconnected by pulling it off the s...

Страница 3: ...out in the same way NOTE Do not discard the transit brackets or screws as these will be needed again if the unit is to be transported from its current location 8 9 From the fittings kit select the 6mm...

Страница 4: ...gle phase power supply NOTE Do not touch anything inside the generator while the side panels are removed and the mains power is connected to the unit 12 13 Switch the generator power ON at the switch...

Страница 5: ...witch the generator power OFF at the switch on the front panel Remove the mains cable completely from the power supply 16 17 Re attach both the RH LH side panels to the generator taking care to ensure...

Страница 6: ...ion Select desired system using highlighted switches 20 21 Switch the generator back ON Leave the generator switched on for a period of 30 minutes to purge the system 22 23 The CURTAIN SOURCE EXHAUST...

Страница 7: ...OGEN ports It is recommended that PTFE tape is applied to the thread to ensure a leak tight fit 26 27 Select the appropriate size Teflon Tubes and connect to the CURTAIN SOURCE EXHAUST NITROGEN ports...

Страница 8: ...d transformer at your earliest convenience Mains voltages between 220V and 253V no transformer is required The 12m length of Teflon tube supplied in the fittings kit is to allow up to 3m of tubing per...

Страница 9: ...ntain Crescent Inchinna Business Park Inchinnan PA4 9RE Scotland UK T 44 0 141 812 8100 F 44 0 141 812 8200 W www peakscientific com E info peakscientific com Part No IG Genius3040 Rev 3 Page Page 9 o...

Отзывы: