Normal Operation

The Genius 3010/3013/3020 & 3022 Gas Generators are designed specifically to 

minimize operator involvement. Given that the systems are installed as described 

in earlier sections and is serviced in accordance with the specified maintenance 
recommendations (see Service Requirements), then it should simply be a matter of 

turning the Generator on when it is required. 

The Generator will automatically produce the factory set flow and pressure as detailed 

in the Technical Specifications.

On Demand Gas

The generator will produce nitrogen on demand. If the application is operating 
and requires a gas flow, the system it is connected to will supply this to suit the 
requirements of the application. If the application requirement for gas stops, the system 
will also stop, once it has reached its upper set limit in the internal storage tanks. If the 
demand from the application starts again, the system will detect the demand for gas 
and will automatically start again to suit the demand.

Generator Cycling

The Generator is designed for the internal compressors to cycle. This cycling reduces 

the duty (run time) on the compressors. The rate at which they cycle will be dependent 
on the gas required to satisfy the demand of the application. If the application demands 

the maximum gas flow of the Generator, the compressor duty will be higher, (the rest 

period in the compressor cycle will be shorter). If the application demand is lower than 
the maximum gas flow, then the duty on the compressors will decrease, (the rest period 

in the compressor cycle will be longer). If the Generator is installed in an extreme 

environment or is subjected to low supply voltage or high altitude the compressors may 

undergo periods where they do not cycle.

Unusual Operation

If at any time the generator begins to emit excessive noise or vibration, then it should 

be switched off and you should contact Peak Scientific or the Peak Partner from which 
the generator has been purchased.

Содержание Genius 3010

Страница 1: ...Genius3010 3013 3020 3022 User Manual Rev 1 PN UM 0008 03 02 17 Genius 3010 13 20 22 User Manual...

Страница 2: ...ngs Kit Contents 14 Installation 15 Generator Environment 15 3010 3013 General Dimensions 16 3010 3013 Rear Connections 17 3020 3022 General Dimensions 18 3020 3022 Rear Connections 19 Electrical Conn...

Страница 3: ...ant information Users can refer to the contents page to find the relevant information Please review each of the following sections carefully Thank you for selecting Peak Scientific to meet your gas ge...

Страница 4: ...Generator designed as a plug and play system Economical More cost effective than any other gas supply method Convenient Gas on demand no hazards no need to worry about running out of gas The Genius 30...

Страница 5: ...ality of the foregoing that any defect is not the direct or indirect result of lack of repair and or servicing incorrect repair and or servicing use of wrong materials and or incorrect spare parts 2 A...

Страница 6: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 7: ...uirements Low Voltage Directive 2015 35 EU EN 61010 1 2010 Electrical Equipment for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical...

Страница 8: ...19 EU These obligations are being met within the B2B compliance group Peak Scientific Instruments Ltd has developed all reasonable due diligence controls to ensure that our products comply with the p...

Страница 9: ...r 100 psi Maximum Outlet Flow Rate 64 L min 2 26 cfm Start Up Time 30 mins Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 1 x BSPP Drain Outlets 1 x BSPP Pressure Gauges 1 Electri...

Страница 10: ...let Flow Rate 64 L min 2 26 cfm Chip Cube Mode increased O2 content to 10 Start Up Time 30 mins Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 1 x BSPP Drain Outlets 1 x BSPP Pres...

Страница 11: ...r 100 psi Maximum Outlet Flow Rate 2 x independant outlet of 32 L min 1 13 cfm Start Up Time 30 mins Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 2 x BSPP Drain Outlets 1 x BSPP...

Страница 12: ...r 116 psi Maximum Outlet Flow Rate 2 x independant outlet of 32 L min 1 13 cfm Start Up Time 30 mins Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 2 x BSPP Drain Outlets 1 x BSPP...

Страница 13: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Страница 14: ...Tubing x 3m 2 Teflon Tubing x 3m 3 6mm PE Tubing x 3m 4 Compression Fitting x 1 5 6mm Push Fit Fitting x 2 6 Flow Control Silencer x 1 7 UK Mains Power Cable x 1 8 EU Mains Power Cable x 1 9 US Mains...

Страница 15: ...s affected by ambient conditions Note should also be taken to the proximity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the uni...

Страница 16: ...16 Generator Overview General Dimensions The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 600 mm 23 6 850 mm 33 5 1322...

Страница 17: ...Page 17 3010 3013 Rear Connections Compression Fitting 6mm Push Fit Fitting Mains Cable...

Страница 18: ...Page 18 3020 3022 Rear Connections Compression Fitting 6mm Push Fit Fitting Mains Cable...

Страница 19: ...s classified as SAFETY CLASS 1 THIS UNIT MUST BE EARTHED Before connecting the unit to the mains supply please check the information on the serial plate The mains supply must be of the stated AC volta...

Страница 20: ...hing the output pressure gauges on the front panel This will climb to the factory set pressure as noted in the specifications Once this pressure is reached the compressors will continue to run until t...

Страница 21: ...cation starts again the system will detect the demand for gas and will automatically start again to suit the demand Generator Cycling The Generator is designed for the internal compressors to cycle Th...

Страница 22: ...e application can lead to a reduction in efficiency Tubing Lengths The diameter of the tubing which will be connected to the gas outlet is important and is determined by the length of tubing required...

Страница 23: ...ed Generators Kit Includes 4 x 0 01 Micron Filter Element 2 x RAC Filter Element 1 x Inlet Filter Element 2 x 3 2 Valve 4 x 2 2 Valve 6 x Plug In Relay 4 x Comp Service Kit 08 9408 1 Genius 3020 22 An...

Страница 24: ...lly for a period of 8 hours There are a number of extreme conditions that can affect the duty cycling of the compressors inside the generator These extremes include high ambient temperatures low suppl...

Страница 25: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 26: ...e cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cle...

Страница 27: ...ng Check pressure gauges are showing normal pressure Contact your service provider The mass spec is reporting low pressure Contact your service provider Yellow SERVICE LED on front panel is on constan...

Страница 28: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 29: ...leading rapid response service to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak to your...