background image

 

Document Number PD196-068 R1         

14 October 2004 

Page 8 of 24 

 

 

     Testing Type 

UL and cUL 

 

     Healthcare Certified 

Yes, UL6500 and cUL6500 

 

Interference 

FCC, CFR 47 Part 15 Class B, DOC Class B  

 X-Radiation 

N/A 

 Manufacturing 

Origin 

United 

States 

 

 

 

5. ON SCREEN DISPLAY 

 

 

Menu Type 

Character  

 

Menu Entrance 

IR Remote Control Only 

 

Format 

15 Rows by 30 Columns 

 

Colors 

15 Foreground, 7 Background 

 Menus 

 

 

       Picture 

 

 

            Brightness 

Yes 

 

            Contrast 

Yes 

 

            Color 

Yes 

 

            Tint 

Yes 

 

      Channel Setup 

 

 

            Signal 

Air, CATV, CATV-IRC, CATV-HRC 

 

            Autoprogram 

 

 

                  Free 

Programmable Tier 

 

                  Basic 

Programmable Tier 

 

                  Premium 

Programmable Tier 

 

            Clear Service Level 

 

 

                  Free  

Erasable Tier 

 

                  Basic 

Erasable Tier 

 

                  Premium 

Erasable Tier 

 

            Add/Delete Channels   

 

                  Free 

Individual Channels 

 

                  Basic 

Individual Channels 

 

                  Premium 

Individual Channels 

 

            Copy Service Levels 

Yes, Tier to Tier Copy 

 

      Features 

 

 

               Power ON Channel & Spkr 

Selectable by Tier 

 

                  Free 

Start Channel, Speaker On/OFF 

 

                  Basic 

Start Channel, Speaker On/OFF 

 

                  Premium 

Start Channel, Speaker On/OFF 

 

      Volume Limiter 

Yes 

 

      Caption Text Mode 

Enabled, Disabled 

 

      Language 

English, French, Spanish 

 

      Service Levels 

Disabled, Free, Basic, Premium 

 

      Paid 

Yes, No 

 

      Self Rent 

Locked, Unlocked 

 

      Channel Number 

Yes 

 

      Channel Labels 

No 

 

      Sleep Timer 

No 

 

      Sound Mute 

Yes 

 

      V-Chip Rating 

No 

 

 

 

5. FEATURES 
 Sleep 

Timer  No 

 On/Off 

Timer  No 

 

Auto Shut OFF 

Yes - Disable Mode, Empty (Blank) Tier 

Содержание P10LCD PERSONA 10

Страница 1: ...ge 1 of 24 Service Manual Communication Systems Inc 40 Greenwood Lane Springboro Ohio 45066 PH 937 743 6010 FX 937 743 5664 http www pdiarm com PERSONA10 Hospital Grade LCD Television Model P10LCD Col...

Страница 2: ...ervice industry of all conceivable methods in which service might be done or of the possible hazardous consequences of each method Accordingly a servicer who uses a service procedure or tool which is...

Страница 3: ...sclaimer and Trademarks 4 Regulatory Information 4 Product Identification 5 Remote Control 6 General Specification 7 Disinfecting and Cleaning 12 Hidden Factory Menu 13 Picture and Power Supply Adjust...

Страница 4: ...trade names of their respective holders PDI and Better Solutions Are Within Reach are registered trademarks of PDI Communication Systems Inc Springboro Ohio Regulatory Information FCC This equipment h...

Страница 5: ...be C and so on 0123 456789 A Communication Systems Inc CAUTION ATTENTION RISK OF ELECTRIC SHOCK DO NOT OPEN RISQUE DE CHOC ELECTRIETUE NE PAS QUVRIR TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT REMOVE...

Страница 6: ...ange to prevent dual programming of an adjacent room TV Stand within 3 feet or closer from the front of the TV to use the remote PROGRAMMING REMOTE PD108 412 SETUP Activates On Screen Programming Menu...

Страница 7: ...tion 10 at 0 8W Earphone Location Side Mounted Size 3 5mm Monaural Yes Stereo No Internal Speaker Disabled during Earphone Use Computer Monitor Capable No Internal DVD No Keylock None TV Disable Avail...

Страница 8: ...ium Programmable Tier Clear Service Level Free Erasable Tier Basic Erasable Tier Premium Erasable Tier Add Delete Channels Free Individual Channels Basic Individual Channels Premium Individual Channel...

Страница 9: ...Hospital Welcome Channel Yes Selectable on per Tier Basis Comb Filter Yes Auto Channel Memory Yes Closed Captions Yes CC1 CC2 TEXT1 TEXT2 Text Mode Disable Yes Speaker Mute Yes 6 KEYPAD CONTROLS Swit...

Страница 10: ...peaker No Power RF Input F Type 7 REMOTE IR Format Custom IR Range Range Limited to1 Meter Power Source AAA x 2 Total Keys 29 Keys Power Yes 1 Yes 2 Yes 3 Yes 4 Yes 5 Yes 6 Yes 7 Yes 8 Yes 9 Yes 0 Yes...

Страница 11: ...None LCD Screen Protector Yes Supplemental Non Scratch Weight Approx 7 lbs Ambient Temperature Operating 15 C to 30 C Storage 20 C to 60 C Relative Humidity 5 to 80 Operational 9 SERVICE Reliability...

Страница 12: ...fecting 1 The P10LCD has been designed to withstand up to a 5 chlorine based disinfectant Many alcohol and ammonia based disinfectants have also been tested with success However before using any clean...

Страница 13: ...lowing IR remote key sequence SETUP SETUP 4 0 2 The TV will turn ON automatically The factory screen is displayed PDI Communications Systems Inc PDI P10LCD LCD Healthcare Television Main Sfw V 1 00 CK...

Страница 14: ...tube display and does not require any internal adjustments Dynamic picture adjustments to Brightness Contrast Color and Tint are available using the IR Remote Control part number PD108 412 using the...

Страница 15: ...Document Number PD196 068 R1 14 October 2004 Page 15 of 24 Pan Adjustment Tighten or Loosen Tension Nut to Adjust PAN Friction Arm Nose...

Страница 16: ...Document Number PD196 068 R1 14 October 2004 Page 16 of 24 Tilt Adjustment 1 Remove Access Cover 2 Adjust Tension Nut to Hold Tilt Position 3 Replace Cover...

Страница 17: ...068 R1 14 October 2004 Page 17 of 24 Main Circuit Board Interconnections J18 J12 J11 J5 J7 LCD Panel Backlight Connector LCD Panel Backlight Connector LCD Panel Data Ribbon Coax Swivel Cable Assy PD10...

Страница 18: ...ument Number PD196 068 R1 14 October 2004 Page 18 of 24 IR Speaker Board Interconnections ESD Ground Ribbon to Metal Sub Plate IR Ribbon to J11 Main PCB Speaker Assy PD106 431 EMI Shield IR Speaker Bo...

Страница 19: ...of the swivel resistance be recorded as part of the service record 3 AC LEAKAGE CURRENT TEST Following re assembly of the P10LCD always perform an AC Leakage Current Test Use either an approved hospit...

Страница 20: ...R C network to the metal ring on the earphone jack f Measure the AC voltage drop across the R C network The measured voltage must not exceed 0 10 volts RMS 100 millivolts RMS This corresponds to 100 m...

Страница 21: ...p Shroud PD133 588 Coax Cable Assy PD106 430 Adjustment Cover PD133 590 Rear Housing PD133 587 Main Chassis Printed Circuit Board PD128 1378 LCD Panel PD289 101 Anti Glare Window Assy PD133 593 Speake...

Страница 22: ...hassis BLACK PICTURE SOUND OK 1 Check for on screen menu and adjust for maximum brightness and contrast 2 Verify J5 ribbon connector is seated 3 Verify 12V peak 62 Khz switching voltage waveform is pr...

Страница 23: ...is set in Setup Channel Setup menu 5 Replace main chassis TV KEYPAD INOPERATIVE 1 Verify TV is correctly powered 2 Check TV operation with remote control 3 Verify J11 Ribbon connector is seated 4 If...