PDi P10LCD PERSONA 10 Скачать руководство пользователя страница 2

 

Document Number PD196-068 R1         

14 October 2004 

Page 2 of 24 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Voltage Range 

AC 18-32 

Volts

 

DC 18-32 

Volts

 

 
 
 
 
 
 
 
 

 

Graphical Symbols 

This lightning flash with arrowhead symbol, within an equilateral is 
intended to alert the user of the presence of uninsulated “dangerous 
voltage” within the product’s enclosure that may be of sufficient 
magnitude to constitute a risk of electric shock to persons. 

 
 

The exclamation point within an equilateral triangle is intended to alert 
the user of the presence of important operating and maintenance 
(servicing) instructions in the literature accompanying the appliance. 

 
This service manual contains various CAUTIONS and WARNINGS 
indicated by triangular warning symbols, which should be read and 
understood in order to minimize the risk of personal injury to service 
personnel and customers. The possibility exits that improper service 
methods may damage the equipment or result in property damage or 
user injury. It also is important to understand that these CAUTIONS and 
WARNINGS are not exhaustive. PDI could not possibly know, evaluate 
and advise the service industry of all conceivable methods in which 
service might be done or of the possible hazardous consequences 
of each method. Accordingly, a servicer who uses a service procedure or 
tool which is not recommended by PDI must first satisfy themselves 
thoroughly that neither their safety nor the safe operation of the 
equipment will be compromised by the service method selected. 
 

Product Safety Servicing Precautions 

1. 

MODIFICATIONS

  Do not attempt to modify this product in any way 

without written authorization from PDI. Unauthorized modifications 
will not only void the warranty, but may lead to your being liable for 
any resulting property damage or user injury. 
 

2. 

POWER SOURCE

  Use only a power source from a CSA Certified / 

UL Approved Class 2 Power Supply suitable for use in a Health Care 
Facility.  This TV will operate on either DC or AC voltage. 
 

3. 

X-RAYS

  X-Ray Radiation is not a concern since this TV does not 

incorporate a CRT. 

 
4. 

REPLACEMENT PARTS

  Parts critical to the safe operation of this 

television are marked with a       on schematics or drawings.  
Replace only with the part number specified. 
 

5. 

SAFETY CHECKS  

This hospital grade television requires special 

safety checks before returning to service.  Observe and follow the 

“Safety Check”

 section in this service manual. 

 

 

Содержание P10LCD PERSONA 10

Страница 1: ...ge 1 of 24 Service Manual Communication Systems Inc 40 Greenwood Lane Springboro Ohio 45066 PH 937 743 6010 FX 937 743 5664 http www pdiarm com PERSONA10 Hospital Grade LCD Television Model P10LCD Col...

Страница 2: ...ervice industry of all conceivable methods in which service might be done or of the possible hazardous consequences of each method Accordingly a servicer who uses a service procedure or tool which is...

Страница 3: ...sclaimer and Trademarks 4 Regulatory Information 4 Product Identification 5 Remote Control 6 General Specification 7 Disinfecting and Cleaning 12 Hidden Factory Menu 13 Picture and Power Supply Adjust...

Страница 4: ...trade names of their respective holders PDI and Better Solutions Are Within Reach are registered trademarks of PDI Communication Systems Inc Springboro Ohio Regulatory Information FCC This equipment h...

Страница 5: ...be C and so on 0123 456789 A Communication Systems Inc CAUTION ATTENTION RISK OF ELECTRIC SHOCK DO NOT OPEN RISQUE DE CHOC ELECTRIETUE NE PAS QUVRIR TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT REMOVE...

Страница 6: ...ange to prevent dual programming of an adjacent room TV Stand within 3 feet or closer from the front of the TV to use the remote PROGRAMMING REMOTE PD108 412 SETUP Activates On Screen Programming Menu...

Страница 7: ...tion 10 at 0 8W Earphone Location Side Mounted Size 3 5mm Monaural Yes Stereo No Internal Speaker Disabled during Earphone Use Computer Monitor Capable No Internal DVD No Keylock None TV Disable Avail...

Страница 8: ...ium Programmable Tier Clear Service Level Free Erasable Tier Basic Erasable Tier Premium Erasable Tier Add Delete Channels Free Individual Channels Basic Individual Channels Premium Individual Channel...

Страница 9: ...Hospital Welcome Channel Yes Selectable on per Tier Basis Comb Filter Yes Auto Channel Memory Yes Closed Captions Yes CC1 CC2 TEXT1 TEXT2 Text Mode Disable Yes Speaker Mute Yes 6 KEYPAD CONTROLS Swit...

Страница 10: ...peaker No Power RF Input F Type 7 REMOTE IR Format Custom IR Range Range Limited to1 Meter Power Source AAA x 2 Total Keys 29 Keys Power Yes 1 Yes 2 Yes 3 Yes 4 Yes 5 Yes 6 Yes 7 Yes 8 Yes 9 Yes 0 Yes...

Страница 11: ...None LCD Screen Protector Yes Supplemental Non Scratch Weight Approx 7 lbs Ambient Temperature Operating 15 C to 30 C Storage 20 C to 60 C Relative Humidity 5 to 80 Operational 9 SERVICE Reliability...

Страница 12: ...fecting 1 The P10LCD has been designed to withstand up to a 5 chlorine based disinfectant Many alcohol and ammonia based disinfectants have also been tested with success However before using any clean...

Страница 13: ...lowing IR remote key sequence SETUP SETUP 4 0 2 The TV will turn ON automatically The factory screen is displayed PDI Communications Systems Inc PDI P10LCD LCD Healthcare Television Main Sfw V 1 00 CK...

Страница 14: ...tube display and does not require any internal adjustments Dynamic picture adjustments to Brightness Contrast Color and Tint are available using the IR Remote Control part number PD108 412 using the...

Страница 15: ...Document Number PD196 068 R1 14 October 2004 Page 15 of 24 Pan Adjustment Tighten or Loosen Tension Nut to Adjust PAN Friction Arm Nose...

Страница 16: ...Document Number PD196 068 R1 14 October 2004 Page 16 of 24 Tilt Adjustment 1 Remove Access Cover 2 Adjust Tension Nut to Hold Tilt Position 3 Replace Cover...

Страница 17: ...068 R1 14 October 2004 Page 17 of 24 Main Circuit Board Interconnections J18 J12 J11 J5 J7 LCD Panel Backlight Connector LCD Panel Backlight Connector LCD Panel Data Ribbon Coax Swivel Cable Assy PD10...

Страница 18: ...ument Number PD196 068 R1 14 October 2004 Page 18 of 24 IR Speaker Board Interconnections ESD Ground Ribbon to Metal Sub Plate IR Ribbon to J11 Main PCB Speaker Assy PD106 431 EMI Shield IR Speaker Bo...

Страница 19: ...of the swivel resistance be recorded as part of the service record 3 AC LEAKAGE CURRENT TEST Following re assembly of the P10LCD always perform an AC Leakage Current Test Use either an approved hospit...

Страница 20: ...R C network to the metal ring on the earphone jack f Measure the AC voltage drop across the R C network The measured voltage must not exceed 0 10 volts RMS 100 millivolts RMS This corresponds to 100 m...

Страница 21: ...p Shroud PD133 588 Coax Cable Assy PD106 430 Adjustment Cover PD133 590 Rear Housing PD133 587 Main Chassis Printed Circuit Board PD128 1378 LCD Panel PD289 101 Anti Glare Window Assy PD133 593 Speake...

Страница 22: ...hassis BLACK PICTURE SOUND OK 1 Check for on screen menu and adjust for maximum brightness and contrast 2 Verify J5 ribbon connector is seated 3 Verify 12V peak 62 Khz switching voltage waveform is pr...

Страница 23: ...is set in Setup Channel Setup menu 5 Replace main chassis TV KEYPAD INOPERATIVE 1 Verify TV is correctly powered 2 Check TV operation with remote control 3 Verify J11 Ribbon connector is seated 4 If...

Страница 24: ...ent Number PD196 068 R1 14 October 2004 Page 24 of 24 Communication Systems Inc 40 Greenwood Lane Springboro Ohio 45066 USA Internet http www pdiarm com email pdi pdiarm com PH 937 743 6010 FX 937 743...