5. Measurements

 

 

The  measurements  should  be  conducted  in  at  least several places.  These places 

should be as flat, smooth and clean as possible. The actual result of the measurements is 
the  arithmetic  average  of  all  the  measurements.  A  line drawn  between  the  electrodes  
should  be  perpendicular  to the  fibres.  If  the  timber  is  too  thin,  place  the  electrodes 
parallel to the fibres. 

Hold the device as shown on fig 2. Press the electrodes hard against the wood. Read the 

result. 

There should not be any metal elements, wet wood or concrete  etc  under  the  timber.  

It’s  best  if  the  timber  is 
supported   only   by   it’s   ends   or   placed   on   a   layer   of 
Styrofoam several centimetres thick. 

The  instrument  is  equipped  with  an  automatic  power off which activates 5 minutes 

after the last change in upper 
control  knob  setting.  It  can  be  switched  off  instantly  by 
setting the upper control knob to OFF. 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

 
 

Fig. 2 

 
 
 

6. Remarks

 

The size of the timber should not be less than 80 x 40 mm. Measuring moisture 

contents of smaller pieces leads to results lower than the real moisture contents. 

The electrodes are made of special silicon bronze and during normal operation should 

not get deformed. If due to excess force or accidental fall the electrodes get deformed, they  
should  be  bend  to  their  original  shape.  Deformed  or misplaced electrodes are the source  
of  errors  and  in  extreme  cases  can  prevent  proper calibration of the device. 

Avoid  measurements  when  the  timber  or  the  operator of  the  device  are  loaded  with  

high  electrostatic  charges. Such charges (sparks) make measurements difficult and can 
damage the instrument. 
 

 
 
 
 

Содержание PCE- WP21

Страница 1: ...029 03 976 99 29 info warensortiment de www warensortiment de PCE Group Ib rica S L Tel 34 967 543 548 Fax 34 967 543 542 info pce iberica es www pce iberica es PCE Group Italia S R L Tel 39 0583 975...

Страница 2: ...cation 3 2 Technical data 3 3 Description of the controls 3 4 Preparing the instrument 4 5 Measurements 5 6 Remarks 5 7 Density of selected wood types 6 8 Calculating wood density 8 9 Errors 9 10 Chan...

Страница 3: ...tion 60 mm Display LCD 12 7 mm Resolution 0 1 Settings range density 0 3 1 1 g cm thickness 10 60 mm Minimal settings step density 0 02 g cm thickness 2 mm Power battery 6F22 9V Battery life approx 50...

Страница 4: ...llow these steps set the upper control knob to DENSITY and then set the required density using the lower control knob set the upper control knob to THICKNESS and set the required thickness with the lo...

Страница 5: ...ped with an automatic power off which activates 5 minutes after the last change in upper control knob setting It can be switched off instantly by setting the upper control knob to OFF Fig 2 6 Remarks...

Страница 6: ...7 Density of selected wood types Californian Redwood 0 39 Mountain Cherry 0 54 Chestnut Sweet 0 55 Douglas Fir Oregon 0 47 Eberesche 0 73 Ebony Macassar 1 Elm 0 64 European Cherry 0 55 Spruce European...

Страница 7: ...5 Oak White 0 74 Okoume Gaboon 0 41 Pear 0 7 Pine insignis 0 4 Pine Weymouth 0 37 Pine Europ Black 0 53 Pine Redwood 0 49 Popular Black Populus nigra 0 41 Ramin 0 56 Rubberwood 0 49 Redoak Northern 0...

Страница 8: ...cm3 V a b h e g 94 2 cm3 b weight the cube with the accuracy of 1 gram note the result G in grams e g 62g c measure the approximate moisture contents with Wood Moisture Meter WIP 22D setting thickness...

Страница 9: ...equipped with a power level control circuit When the power drops below acceptable level a sign LO BAT appears on the left side of the LCD This indicates that the battery has expired and should be rep...

Отзывы: