9  Troubleshooting 

 

9.1  Faults 

 
Fault 

Possible cause 

Remedy 

It is not possible to power  
on the fan anemometer  
 

On/Off key pressed too  
fleetingly  

Press the On/Off key slightly 
longer  

Batteries not inserted  
correctly  

Check correct polarity 

Battery voltage insufficient 

Replace batteries 

 

9.2  Error codes 

 

9.2.1  Air temperature 

 
Error code 

Error 

Remedial action 

E02  

Temperature is below measuring 
range  

Acclimatize the fan anemometer 
at room temperature for 30 
minutes. If E02 is still displayed, 
return the fan anemometer for 
repair. 

E03  

Temperature is above measuring 
range  

Acclimatize the fan anemometer 
at room temperature for 30 
minutes. If E02 is still displayed, 
return the fan anemometer for 
repair. 

E31  

Temperature sensor defective 

Return the fan anemometer for 
repair. 

 

9.2.2  Relative humidity 

 
Error code 

Error 

Remedial action 

E04 

Caused by temperature error  

Please refer to the errors of air 
temperature 

E11 

Calibration error 

Return the fan anemometer for 
relative humidity calibration  

E31  

Humidity sensor defective 

Return the fan anemometer for 
repair. 

 

9.2.3  Dew point and wet-bulb temperature 

 
Error code 

Error 

Remedial action 

E04 

Caused by temperature or 
humidity error  

Please refer to the paragraphs 
above 

 
 

 

Содержание PCE-VA 20

Страница 1: ...ruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce i...

Страница 2: ...unctions 7 6 1 Data Hold 7 6 2 MIN MAX 7 6 3 Averaging 8 6 3 1 Multi point average 8 6 3 2 Time average 8 6 4 Auto power off function 8 7 SETUP 9 7 1 1 P1 0 Unit selection 9 7 1 2 P2 0 AREA Size 9 7 1...

Страница 3: ...be used by qualified personnel and repaired by PCE Instruments personnel There is no warranty of damages or injuries caused by non observance of the manual Don t use the device with wet hands Do not u...

Страница 4: ...measured value 0 3 m s when above 20 m s Temperature 0 6 C Relative humidity 3 at 25 C else 5 Technical Data Fan diameter 10 cm Fan anemometer dimensions 269 x 160 x 51 mm Airflow cone dimensions Rou...

Страница 5: ...ure dew point temperature 4 1 Display Display indication Meaning HOLD Freeze the current display AVG Averaging MAX MIN Highest and lowest values Low battery warning m s fpm Flow velocity display units...

Страница 6: ...MIN and MAX values In average mode press to record data or start timed recording In setup mode press to enter data settings Switches to upper display Long press to go into average mode In average mode...

Страница 7: ...ttach one of the optional cones to the meter The anemometer detects the cone automatically and shows a cone indication on the display After that the device will calculate the volume flow Note Make sur...

Страница 8: ...e current reading Press the Hold button to start recording The timer on the upper display starts counting Press the Up button to stop and calculate the timed average You can see the duration on the up...

Страница 9: ...nsions of the cross sectional area for volume flow measurements Press the Hold button to enter the area setting The lower display now shows 99999 with the first digit flashing Press the Up button to c...

Страница 10: ...fan anemometer for repair E03 Temperature is above measuring range Acclimatize the fan anemometer at room temperature for 30 minutes If E02 is still displayed return the fan anemometer for repair E31...

Страница 11: ...s above range Take measurements within range If E03 is still displayed return the fan anemometer for repair 9 2 5 Volume Flow Error code Error Remedial action E03 Value is above display range Check en...

Страница 12: ...contact PCE Instruments 10 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 10 2...

Отзывы: