Manual
18
Assembly
Remove the nut and the washer (1) from the rack (2).
Using the nut and the washer (1), fix the extension rod (3)
between the stand and the rack (2). The extension rod is
needed due to the length of the LCP adapters. Without the
extension rod, it would be difficult to attach the adapter to the
viscometer.
Using the screw (6), attach the water jacket (4) onto the rear
support (5).
Close the sample container (7) with the lower cap (8). Ensure
that it is sufficiently tight.
Fill the sample container (7). Be careful that there are not any
bubbles in the container. Use a large syringe to fill the
container. The sample amount needed is very small (18 ml).
Check the filling level. The spindle should be completely
covered to be the correct sample amount.
Hang the LCP spindle (9) onto the spindle hook (10). Attach the
hook to the screw (11).
Insert the spindle with the hook (10) and screw (11) into the
sample container.
Insert the sample container (7) into the water jacket (4) from
below.
Fix the sample container (7) in the water jacket (4). The groove
of the lower cap (12) has to be aligned with the fastening screw
(13) of the water jacket. Turn the sample container to do so.
Attach the upper cap
(14). Make sure the spindle does not fall
into the sample container. Let the spindle lean on the upper
cap.
Using the provided screw (15), attach the rear support (5) to the
viscometer.