background image

 

 

©  PCE Instruments 

15 

 

To set the alarm limits, press the 

 key in the current measuring mode. 

 

 

 

Display 

Meaning 

 

This setting refers to tensile force 

 

This setting refers to compressive force 

L. Limit 

Lower Limit. Here you set the first limit value. This value cannot be higher than 
the H. Limit value. 

H. Limit 

Higher Limit. Here you set the second limit value. This value cannot be smaller 
than the L. Limit value. 

Number 

Here you set the desired limit.  

 

Use the arrow keys to select the desired parameter. Now press the 

 key to make 

changes to this value. You can now use the arrow keys to change the value as desired. Confirm 

the entry with the 

 key. Press the

 key to return to the measuring mode. 

 

Note:

 

The second limit value must always be higher than the first set limit value. The set values are 

shown above the display in measuring mode. Pressing the 

 key selects between the 

display of the set alarm tension or the set alarm compression. 
 

 

Содержание PCE-PFG 100

Страница 1: ...es Force Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...ns 4 6 Battery 5 7 Switching on and off 5 8 Measuring modes 6 8 1 Real Time 6 8 2 Peak 6 8 3 Average mode 6 8 4 Save mode Automatic memory 12 9 Alarm limits 14 10 Rotate display 16 11 Communication in...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...2 85 V when active Protection class IP 54 Operating and storage conditions 5 45 C 35 65 RH non condensing Dimensions 189 x 707 x 34 mm Weight 450 g Models with external load cell Dimensions weight lo...

Страница 5: ...PCE Instruments 3 4 Dimensions Dimensions for models with internal measuring cell Dimensions for models with external measuring cell...

Страница 6: ...e unit in measuring mode Increasing displayed parameter Selecting the next item in the menu Enter key to confirm the entry or to open a menu item DIR key for switching between compressive and tensile...

Страница 7: ...wing icon is displayed When the battery is fully charged the icon appears on the display When switched on the battery indicator shows the current battery capacity when no charger is connected Battery...

Страница 8: ...thing the range of 10 of the total capacity 8 1 Real Time In Real Time RT measuring mode the current measured value is continuously displayed 8 2 Peak In peak mode PK the highest measured value is dis...

Страница 9: ...suring procedure is possible over a period of 10 minutes As long as the measurement time of 10 minutes is not exceeded this measurement can be resumed at any time To make settings in this measuring mo...

Страница 10: ...0 300 0 s Resolution 0 1 s This parameter only affects the MOD1 function Cap Time Capture Time Here you set the measuring time for the average measurement Available settings 0 0 300 0 s Resolution 0 1...

Страница 11: ...rage measurement and delete individual saved measured values View Ave Stat An average evaluation of all average measurements is carried out here Send Ave Date Here the saved measurement data are trans...

Страница 12: ...delete the selected value In the summary of the average readings you will see the highest the lowest the number of readings and the overall average Under Send Ave Data all data are sent to the PC Aft...

Страница 13: ...he display the force gauge will wait until the set minimum time has elapsed If the minimum load is present and the minimum time has elapsed the actual measurement begins GATHER appears on the display...

Страница 14: ...ction is specified in the settings under P M C F To evaluate the data press the key Then press to select PEAK The following options are now available Selection Description View Peak Data Read out the...

Страница 15: ...arrow keys Use the key to switch between the individual pages Use the key to delete the selected value The peak summary shows you the highest the loweest the number of readings and the total peak Und...

Страница 16: ...limits can be set here If the measured value is lower than the set Lower Limit this is signalled by the red and green LEDs lighting up If the measured value is between the set Higher Limit and Lower...

Страница 17: ...annot be smaller than the L Limit value Number Here you set the desired limit Use the arrow keys to select the desired parameter Now press the key to make changes to this value You can now use the arr...

Страница 18: ...activated with this setting After installing the drivers and the software you can establish a serial connection to the force gauge via the software To do so set the parameters of the force gauge unde...

Страница 19: ...o the meter press the key three times in measuring mode Use the arrow keys to select your desired parameter Once you have found your desired parameter press the key to select the function Use the arro...

Страница 20: ...he end point at which data recording is to be stopped The key must be pressed again for the following functions Function Meaning P M C F Here you set the minimum load for the measuring mode SAVE MODE...

Страница 21: ...06 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that pur...

Страница 22: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Отзывы: